K2M MESA DEFORMITY

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-04-04 for K2M MESA DEFORMITY manufactured by K2m.

Event Text Entries

[104565580] Patient underwent a t10 to s2a1 instrumentation with k2m mesa instrumentation. At surgery all screws were final tightened according to manufacturer specifics. At 6 months, the rods have lost connection to the s2a1 screws at the distal end of the construct. There has been no rod failure. All hardware was in perfect position. At follow up intervals the rods remained connected until 3 months after surgery. This patient had good bone quality and underwent a scoliosis correction with internal rod reduction techniques. Eoc screws were used at s1 and t10. Rods were final tightened at all levels according to k2m specifics. The locking mechanism for the k2m mesa screw may be deficient. I have another case with the exact same problem i would be happy to share. It is my intent that the biomechanics of the final tightening mechanism be evaluated on long construct fusions in the lumbar spine.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5076322
MDR Report Key7398769
Date Received2018-04-04
Date of Report2018-04-03
Date of Event2018-03-23
Date Added to Maude2018-04-05
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameK2M MESA DEFORMITY
Generic NameLUMBAR PEDICLE SCREWS
Product CodeOSH
Date Received2018-04-04
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrI
Device Sequence No1
Device Event Key0
ManufacturerK2M
Manufacturer AddressLEESBURG VA 20175 US 20175


Patients

Patient NumberTreatmentOutcomeDate
101. Other; 2. Required No Informationntervention 2018-04-04