CAVIWAVE PRO ULTRASONIC CLEANING SYSTEM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-04-06 for CAVIWAVE PRO ULTRASONIC CLEANING SYSTEM manufactured by Steris Corporation - Distribution Center.

Event Text Entries

[104833117] All instruments present were manually washed prior to terminal cleaning. A steris service technician arrived onsite to inspect the unit. The technician initiated a test cycle and was able to recreate the reported event. The technician inspected the unit and found that the water solenoid valve was damaged. The steris service technician replaced the solenoid valve, ran a test cycle and confirmed the unit to be operating properly. The caviwave was returned to service. A review of complaint records from the past five years indicate this to be an isolated event. No additional issues have been reported.
Patient Sequence No: 1, Text Type: N, H10


[104833118] The user facility reported that smoke was emitting from their caviwave pro ultrasonic cleaner. No report of injury, procedure delay or cancellation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003950207-2018-00001
MDR Report Key7402342
Date Received2018-04-06
Date of Report2018-04-06
Date of Event2018-03-09
Date Mfgr Received2018-03-09
Date Added to Maude2018-04-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. DANIEL DAVY
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927453
Manufacturer G1STERIS CORPORATION - DISTRIBUTION CENTER
Manufacturer Street6100 HEISLEY RD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal Code44060
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCAVIWAVE PRO ULTRASONIC CLEANING SYSTEM
Generic NameCLEANING SYSTEM
Product CodeFLG
Date Received2018-04-06
OperatorOTHER HEALTH CARE PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTERIS CORPORATION - DISTRIBUTION CENTER
Manufacturer Address6100 HEISLEY RD MENTOR OH 44060 US 44060


Patients

Patient NumberTreatmentOutcomeDate
10 2018-04-06

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.