SMART MONITOR 2PS 1030271

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-04-06 for SMART MONITOR 2PS 1030271 manufactured by Circadiance, Llc.

Event Text Entries

[104845314] The nurse call jack in this monitor was physically damaged rendering the output inoperable. Circadiance has recently re-evaluated reporting decisions related to physically damaged nurse call jacks. Circadiance has made the decision to report these instances of damaged nurse call jacks where the jack is still physically intact and it is possible that the user may have perceived that they were still able to make a connection to the jack. This complaint was received prior to the re-evaluation of reporting, however, the event occurred relatively recently. Thus circadiance has re-classified this complaint and has retrospectively filed this report to be consistent with the recent internal changes regarding reporting these events. The primary alarm in this monitor as well as all of the monitoring parameters were confirmed to be operating according to specification.
Patient Sequence No: 1, Text Type: N, H10


[104845315] (b)(6) hospital sent a smart monitor 2ps model number 1030271, serial number (b)(4) in for service requesting that circadiance check the overall performance of the monitor. The customer did not allege that the product was malfunctioning. They just requested that the device be evaluated. During service evaluation, the service technician determined that the nurse call jack was partially dislodged from the main circuit board rendering the nurse call output nonfunctional.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3006182632-2018-00009
MDR Report Key7402953
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-04-06
Date of Report2018-03-28
Date of Event2018-02-02
Date Mfgr Received2018-01-30
Device Manufacturer Date2009-10-16
Date Added to Maude2018-04-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. RICHARD LORDO
Manufacturer Street1300 RODI ROAD
Manufacturer CityTURTLE CREEK PA 15145
Manufacturer CountryUS
Manufacturer Postal15145
Manufacturer Phone7243879836
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSMART MONITOR 2PS
Generic NameINFANT APNEA MONITOR
Product CodeNPF
Date Received2018-04-06
Returned To Mfg2018-01-30
Model Number1030271
Catalog Number1030271
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCIRCADIANCE, LLC
Manufacturer Address1300 RODI RD. TURTLE CREEK PA 15145 US 15145


Patients

Patient NumberTreatmentOutcomeDate
10 2018-04-06

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.