SUPPORT ARM 177 61481720

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-04-06 for SUPPORT ARM 177 61481720 manufactured by Maquet Critical Care Ab.

Event Text Entries

[104745861] No parts were returned. The broken support arm was discarded by the user facility. Investigation was based on picture received. According to the picture, the support arm broke at the joint nearest to the bracket. The support arm is a casting and it most probably developed a crack at an earlier occasion either by overloading or impact and this led to the reported breaking. The support arm has been successfully tested for mechanical strength and is designed according to standard. It is unknown under which circumstances the support arm broke. Previous investigations of similar faults led to a redesign and change of the manufacturing process in order to obtain a higher mechanical strength of the support arm. The support arm associated with this complaint had been manufactured before this change was implemented in production. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[104745862] It was reported that the support arm broke at the lower end. There was no patient involvement. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8010042-2018-00176
MDR Report Key7403153
Date Received2018-04-06
Date of Report2018-04-06
Date Mfgr Received2018-03-15
Date Added to Maude2018-04-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer StreetMAQUET CRITICAL CARE AB R
Manufacturer CitySOLNA
Manufacturer CountrySW
Manufacturer G1MAGNUS LINDQVIST
Manufacturer StreetMAQUET CRITICAL CARE AB R
Manufacturer CitySOLNA
Manufacturer CountrySW
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameSUPPORT ARM 177
Generic NameSUPPORT, ARM
Product CodeIOY
Date Received2018-04-06
Catalog Number61481720
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerMAQUET CRITICAL CARE AB
Manufacturer AddressSOLNA SW


Patients

Patient NumberTreatmentOutcomeDate
10 2018-04-06

© 2026 FDA.report