PARENCHYMAL BOLT PRESSURE MONITORING KIT * NL950-P

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2006-07-19 for PARENCHYMAL BOLT PRESSURE MONITORING KIT * NL950-P manufactured by Integra Neuro Sciences.

Event Text Entries

[490303] A male pt arrived at the icu in 2006, for severe cranial traumatism management, following an accident in the st. Upon arrival, he presented a glasgow score of 5, that triggered an orotracheal intubation. Cerebral scanner showed a major cerebral edema with suspicion of right carotid dissection. An icp transducer was placed, icp values were paradoxically low, below 15mmhg. A cerebral scanner for control was performed and showed the right transducer position in intraparenchymal tissue. Over the following hours, the displayed icp values became negative so around 12 hours after the first trnasducer was placed, the decision was made to replace it. For the second procedure, a new catheter was used with the same monitor, and the icp value was 50mmhg. The facility's hypothesis at the time was that a calibration error was made prior to placement of the first transducer, even though the same operators performed both procedures.
Patient Sequence No: 1, Text Type: D, B5


[7849034] An investigation has been initiated based on the reported info.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2023988-2006-00032
MDR Report Key740338
Report Source01,05,06
Date Received2006-07-19
Date of Report2006-07-18
Date of Event2006-04-01
Date Mfgr Received2006-06-16
Date Added to Maude2006-07-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationPHARMACIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactLINDA GOODSTEIN
Manufacturer Street311 ENTERPRISE DRIVE
Manufacturer CityPLAINSBORO NJ 08536
Manufacturer CountryUS
Manufacturer Postal08536
Manufacturer Phone6099362269
Manufacturer G1INTEGRA NEUROSCIENCES
Manufacturer Street5955 PACIFIC CENTER BLVD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal Code92121
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePARENCHYMAL BOLT PRESSURE MONITORING KIT
Generic NameICP DISPOSABLES
Product CodeHCA
Date Received2006-07-19
Model Number*
Catalog NumberNL950-P
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Implant FlagY
Date Removed*
Device Sequence No1
Device Event Key728210
ManufacturerINTEGRA NEURO SCIENCES
Manufacturer Address* SAN DIEGO CA * US


Patients

Patient NumberTreatmentOutcomeDate
10 2006-07-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.