MAUDE MDR 7403791

MDR report key
7403791
Report number
3010326005-2018-00030
Event key
0
Event type
3
Date of event
2018-03-13
Date received
2018-04-06
Adverse event
3
Product problem
3
Patients in event
0
Reporter occupation
0
Health professional
3
Initial report to FDA
3
Event location
3

Manufacturer Contact#

Contact
MR TODD SPERLING
Address
5101 SHADY OAK ROAD MINNETONKA MN 55343 US
Phone
763-763-7632
Report source
M
Manufacturer link flag
Y

Devices#

Seq, Brand, Generic table
SeqBrandGenericManufacturerProduct codeModelCatalogLotPMA510(k)ImplantEvaluatedAvailability
1IMRIS ORT300 OPERATING ROOM TABLEOPERATING ROOM TABLEIMRIS - DEERFIELD IMAGING, INC.KXJORT300114093-000N/AY Y

Patients#

Sequence, Received, Treatment table
SequenceReceivedTreatmentOutcome
12018-04-060

Event Narratives#

N

Patient 1

THE TABLE DRIFT WAS RESOLVED BY REPLACING THE CHECK VALVE BLOCK IN THE MANIFOLD. THE TABLE WAS TESTED TO VERIFY PROPER FUNCTION.

D

Patient 1

THE CUSTOMER STATED THAT THE ORT300 TABLE IS DRIFTING IN THE TRENDELENBURG OVERNIGHT WHEN NOT IN USE WHILE UNLOADED. OBSERVED DURING DAILY QA/ INSPECTION.