Patient 1
THE TABLE DRIFT WAS RESOLVED BY REPLACING THE CHECK VALVE BLOCK IN THE MANIFOLD. THE TABLE WAS TESTED TO VERIFY PROPER FUNCTION.
| Seq | Brand | Generic | Manufacturer | Product code | Model | Catalog | Lot | PMA | 510(k) | Implant | Evaluated | Availability |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 1 | IMRIS ORT300 OPERATING ROOM TABLE | OPERATING ROOM TABLE | IMRIS - DEERFIELD IMAGING, INC. | KXJ | ORT300 | 114093-000 | N/A | Y | Y |
| Sequence | Received | Treatment | Outcome |
|---|---|---|---|
| 1 | 2018-04-06 | 0 |
Patient 1
THE TABLE DRIFT WAS RESOLVED BY REPLACING THE CHECK VALVE BLOCK IN THE MANIFOLD. THE TABLE WAS TESTED TO VERIFY PROPER FUNCTION.
Patient 1
THE CUSTOMER STATED THAT THE ORT300 TABLE IS DRIFTING IN THE TRENDELENBURG OVERNIGHT WHEN NOT IN USE WHILE UNLOADED. OBSERVED DURING DAILY QA/ INSPECTION.