SONIC GENERIC ULTRASOUND TRANSMISSION GEL UNK REF P11111-05

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2006-07-21 for SONIC GENERIC ULTRASOUND TRANSMISSION GEL UNK REF P11111-05 manufactured by Pharmaceutical Innovations, Inc.

Event Text Entries

[480278] Infection control noted that nicu babies with positive cultures for pseudomonas had all gone for echocardiograms. Investigation lead to culturing the gel. Ultrasound gel comes in 5 liter containers with 8 oz refillable bottles. Bottles are topped off and not capped. 8 oz bottles and 5 liter bottles tested positive for the pseudomonas organism. A new unopened container was sent out for culture - results are pending. Four areas of the hospital use the refillable bottles, the rest of the hospital buys the 8 oz bottle. Infection control discourages topping off any fluids. It is not known for sure if the ultrasound gel was the source of the nicu babies' infections as it is a common organism. Investigation and testing is ongoing.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number740449
MDR Report Key740449
Date Received2006-07-21
Date of Report2006-07-21
Date of Event2006-07-21
Report Date2006-07-21
Date Reported to FDA2006-07-21
Date Added to Maude2006-07-27
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameSONIC GENERIC ULTRASOUND TRANSMISSION GEL
Generic NameGEL, ULTRASOUND
Product CodeMUI
Date Received2006-07-21
Model NumberUNK
Catalog NumberREF P11111-05
Lot Number122163
ID Number*
Device AvailabilityY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key728318
ManufacturerPHARMACEUTICAL INNOVATIONS, INC
Manufacturer Address897 FRELINGHUYSEN AVE NEWARK NJ 071142195 US

Device Sequence Number: 2

Brand NameSONIC GENERIC ULTRASOUND TRANSMISSION GEL
Generic NameGEL, ULTRASOUND
Product CodeMUI
Date Received2006-07-21
Model NumberUNK
Catalog NumberREF P11111-05
Lot Number122105
ID Number*
Implant FlagN
Date Removed*
Device Sequence No2
Device Event Key728319
ManufacturerPHARMACEUTICAL INNOVATIONS, INC.
Manufacturer Address* * *


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Other; 3. Required No Informationntervention 2006-07-21

© 2026 FDA.report