MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2006-07-21 for SONIC GENERIC ULTRASOUND TRANSMISSION GEL UNK REF P11111-05 manufactured by Pharmaceutical Innovations, Inc.
[480278]
Infection control noted that nicu babies with positive cultures for pseudomonas had all gone for echocardiograms. Investigation lead to culturing the gel. Ultrasound gel comes in 5 liter containers with 8 oz refillable bottles. Bottles are topped off and not capped. 8 oz bottles and 5 liter bottles tested positive for the pseudomonas organism. A new unopened container was sent out for culture - results are pending. Four areas of the hospital use the refillable bottles, the rest of the hospital buys the 8 oz bottle. Infection control discourages topping off any fluids. It is not known for sure if the ultrasound gel was the source of the nicu babies' infections as it is a common organism. Investigation and testing is ongoing.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 740449 |
MDR Report Key | 740449 |
Date Received | 2006-07-21 |
Date of Report | 2006-07-21 |
Date of Event | 2006-07-21 |
Report Date | 2006-07-21 |
Date Reported to FDA | 2006-07-21 |
Date Added to Maude | 2006-07-27 |
Event Key | 0 |
Report Source Code | User Facility report |
Manufacturer Link | N |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 0 |
Reporter Occupation | RISK MANAGER |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Single Use | 0 |
Previous Use Code | 0 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | SONIC GENERIC ULTRASOUND TRANSMISSION GEL |
Generic Name | GEL, ULTRASOUND |
Product Code | MUI |
Date Received | 2006-07-21 |
Model Number | UNK |
Catalog Number | REF P11111-05 |
Lot Number | 122163 |
ID Number | * |
Device Availability | Y |
Implant Flag | N |
Date Removed | * |
Device Sequence No | 1 |
Device Event Key | 728318 |
Manufacturer | PHARMACEUTICAL INNOVATIONS, INC |
Manufacturer Address | 897 FRELINGHUYSEN AVE NEWARK NJ 071142195 US |
Brand Name | SONIC GENERIC ULTRASOUND TRANSMISSION GEL |
Generic Name | GEL, ULTRASOUND |
Product Code | MUI |
Date Received | 2006-07-21 |
Model Number | UNK |
Catalog Number | REF P11111-05 |
Lot Number | 122105 |
ID Number | * |
Implant Flag | N |
Date Removed | * |
Device Sequence No | 2 |
Device Event Key | 728319 |
Manufacturer | PHARMACEUTICAL INNOVATIONS, INC. |
Manufacturer Address | * * * |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Hospitalization; 2. Other; 3. Required No Informationntervention | 2006-07-21 |