HUDSON MASK,NON-REBREATH W/SAFETY VENT,ELONGAT 1059

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2018-04-09 for HUDSON MASK,NON-REBREATH W/SAFETY VENT,ELONGAT 1059 manufactured by Teleflex Medical.

Event Text Entries

[104876015] (b)(4). A visual, dimensional, and functional inspection of the device involved in the complaint could not be conducted since the device or a picture of the alleged defect was not provided at the time of this report. However material from the production line was verified and no issues were found that can lead to this customer complaint. The device history record (dhr) has been reviewed and no issues or discrepancies were found which could potentially be related to this complaint. Dhr shows that the product was assembled and inspected according to our specifications. Customer complaint cannot be confirmed based only on the information provided. To perform a proper investigation to confirm the alleged defect and determine a root cause, it is necessary to evaluate the sample involved. If the device becomes available this report will be updated with the evaluation results.
Patient Sequence No: 1, Text Type: N, H10


[104876016] Customer complaint alleges "the mask works improperly. Leak found during use, at the level of the tube. " the device was changed for a new one. There was no report of patient impact or consequence. Patient condition was reported as "fine".
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004365956-2018-00109
MDR Report Key7407780
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2018-04-09
Date of Report2018-03-28
Date of Event2018-03-27
Date Mfgr Received2018-05-07
Device Manufacturer Date2013-11-27
Date Added to Maude2018-04-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKATHARINE TARPLEY
Manufacturer Street3015 CARRINGTON MILL BLVD
Manufacturer CityMORRISVILLE NC 27560
Manufacturer CountryUS
Manufacturer Postal27560
Manufacturer Phone9194334854
Manufacturer G1TELEFLEX MEDICAL
Manufacturer StreetPARQUE INDUSTRIAL FINSA
Manufacturer CityNUEVO LAREDO 88275
Manufacturer CountryMX
Manufacturer Postal Code88275
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHUDSON MASK,NON-REBREATH W/SAFETY VENT,ELONGAT
Generic NameMASK, OXYGEN, NON-REBREATHING
Product CodeKGB
Date Received2018-04-09
Returned To Mfg2018-04-13
Catalog Number1059
Lot Number02L1302800
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTELEFLEX MEDICAL
Manufacturer AddressRESEARCH TRIANGLE PARK NC


Patients

Patient NumberTreatmentOutcomeDate
10 2018-04-09

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.