[41416] Table top was becoming increasingly difficult to move longitudinally. A long scrape was evident just under the table top that paralleled the table range of travel. Table supports a vision tester, auto-refractometer and a non-contact keratometer. Inspection found the table top rail roller surface had bent from a 90 degree angle to about 120 degree. The difficulty in table top travel and the scrape were a result of the bent rail. With the rail removed, a three foot groove was worn longitudinally inside the aluminum rail. Most of the equipment weight was distributed towards the pt's side of the table. This imbalance put constant upward pressure on two steel rollers on which the rail traversed, one of which over time wore the groove into the rail. The rail wore to the point that it eventually began to bend. If not identified in time there is a potential for the table top to tip over, dumping the equipment to the ground or worse into a pt's lap.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW4001795
MDR Report Key74123
Date Received1997-03-06
Date of Report1996-03-27
Date Added to Maude1997-03-10
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA0
Report to FDA0
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameZEE
Generic NameOPHTHALMIC TRANSLATING TABLE
Product CodeHRJ
Date Received1997-03-06
Model NumberTS-1
Catalog Number72-2804-1
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device Availability*
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key73953
ManufacturerMARCO OPTHALMIC. INC.
Manufacturer Address11825 CENTRAL PKWY JACKSONVILLE FL 32224 US


Patients

Patient NumberTreatmentOutcomeDate
10 1997-03-06