[105518611] The customer contacted siemens customer care center (ccc) and reported that an operator misinterpreted the "no coagulation" error as the "flat curve" error and incorrectly reported a flagged activated partial thromboplastin time (aptt) result of <20 seconds. Siemens investigated the issue and determined that quality controls were within expected ranges at the time of the event and correct results were produced by the sysmex cs-2500 systems. Based on the sysmex cs-2500 evaluation and algorithm bulletin, "no coagulation" error is triggered when the system is unable to detect a coagulation reaction or a weak clot was formed. The user should: "check the sample for possible anticoagulant contamination, hemolysis, lipemia, etc; verify delivery of sample and reagent; reanalyze the sample with an extended measurement time (within the range of 100 to 1800 seconds) or set "measurement time (sub)" in the test protocol; if reanalysis with extended time gives "no coagulation" message again, the sample is below the detection limits of the detector. Make a comprehensive judgment, taking sample and reagent, etc. Into consideration. If the result is plausible it may be reported as "greater than" per the established laboratory policy. If additional verification is indicated, use your laboratory policy. If additional verification is indicated, use your laboratory's policy to confirm the data. " siemens determined that the customer did not follow the instructions in the sysmex cs-2500 evaluation and algorithm bulletin. The cause of the discordant aptt result reported on a patient sample is a use error. The reagent and systems are performing according to specifications. No further evaluation of these devices is required.
Patient Sequence No: 1, Text Type: N, H10


[105518612] Non-numerical activated partial thromboplastin time (aptt) results (flagged with "no coagulation") were obtained on three samples from the same patient using the dade actin fsl activated ptt reagent on two sysmex cs-2500 systems. The lab protocol allows the operator to report results flagged with "no coagulation" as greater than the reporting limit as long as the operator reviews the curve and reference the time at 50% as the reportable result. The final result was reported to the physician(s) as <20 seconds and based on this result, the physician(s) restarted the patient's heparin drip at 2 am on (b)(6) 2018. The lab supervisor was made aware of the situation and notified the physician(s), who stopped the patient's heparin drip at 6 am on (b)(6) 2018. On the same day, the patient's blood was redrawn twice and run using the same reagent. The result obtained on the second redrawn sample was reported to the physician(s). There are no known reports of adverse health consequences due to the flagged aptt result reported on the patient sample.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610806-2018-00041
MDR Report Key7415170
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-04-10
Date of Report2018-05-16
Date of Event2018-03-15
Date Mfgr Received2018-04-20
Date Added to Maude2018-04-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCHRISTINA LAM
Manufacturer Street511 BENEDICT AVE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145243504
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS PRODUCT GMBH
Manufacturer StreetEMIL-VON-BEHRING-STR. 76
Manufacturer CityMARBURG, D-35041
Manufacturer CountryGM
Manufacturer Postal CodeD-35041
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDADE ACTIN FSL ACTIVATED PTT REAGENT
Generic NameDADE ACTIN FSL ACTIVATED PTT REAGENT
Product CodeGGW
Date Received2018-04-10
Model NumberDADE ACTIN FSL ACTIVATED PTT REAGENT
Catalog Number10445714
Lot Number556916
Device Expiration Date2019-03-28
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS PRODUCT GMBH
Manufacturer AddressEMIL-VON-BEHRING-STR. 76 MARBURG, D-35041 GM D-35041


Patients

Patient NumberTreatmentOutcomeDate
10 2018-04-10

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.