FLEXIBLE VIDEO CYSTOSCOPE 11272VNU

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-04-10 for FLEXIBLE VIDEO CYSTOSCOPE 11272VNU manufactured by Karl Storz Endovision.

Event Text Entries

[105141387] The product evaluation concluded the following: there is debris on distal lens and distal tip of the scope, angle cover is cracked and has debris on it. The shaft marker rings are discolored. A nick was found on the handle housing. The evaluation results suggest that user error can be linked to the physical condition of the instrument. Investigation still in progress.
Patient Sequence No: 1, Text Type: N, H10


[105141388] Allegedly, the physician performed a cystoscopy procedure on patient. Patient's urine was cultured and pseudomonas aeruginosa infection was confirmed from the lab results. Patient was treated with antibiotics from a doctor outside of the clinic. The doctor reported that patient is doing fine.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1221826-2018-00027
MDR Report Key7418116
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-04-10
Date of Report2018-03-12
Date of Event2018-02-06
Date Mfgr Received2018-03-12
Device Manufacturer Date2011-01-11
Date Added to Maude2018-04-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL ASSISTANT
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. SUSIE CHEN
Manufacturer Street2151 E. GRAND AVENUE
Manufacturer CityEL SEGUNDO CA 902455017
Manufacturer CountryUS
Manufacturer Postal902455017
Manufacturer Phone4242188100
Manufacturer G1KARL STORZ ENDOVISION
Manufacturer Street91 CARPENTER HILL ROAD
Manufacturer CityCHARLTON MA 01507
Manufacturer CountryUS
Manufacturer Postal Code01507
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameFLEXIBLE VIDEO CYSTOSCOPE
Generic NameFLEXIBLE VIDEO CYSTOSCOPE
Product CodeFBO
Date Received2018-04-10
Returned To Mfg2018-04-02
Model Number11272VNU
Catalog Number11272VNU
OperatorPHYSICIAN
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerKARL STORZ ENDOVISION
Manufacturer Address91 CARPENTER HILL ROAD CHARLTON MA 01507 US 01507


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-04-10

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.