ROTOPRONE 209500

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2018-04-12 for ROTOPRONE 209500 manufactured by Arjohuntleigh, Inc..

Event Text Entries

[105894791] Please note that this product is no longer manufactured and previous medwatch reports for this product may have been submitted for the manufacturing site kinetic concept inc (under registration #(b)(4)). From (b)(6) 2012 until 2014 complaints related to this product were handled by arjohuntleigh inc, and any medwatch reports were submitted under registration # (b)(4). From 2014 and going forward complaints related to these products are to be handled by arjohuntleigh ab's complaint handling establishment and any medwatch reports will be submitted under registration # (b)(4). Additional information will be provided upon conclusion of the investigation.
Patient Sequence No: 1, Text Type: N, H10


[105894792] A customer called on 24-hour phone line reporting several issues: they could not lock the head support piece from going up and down, the buckle was stuck all day, the hatch was hard to close. There was no injury reported. Arjohuntleigh clinical consultant provided guidance and explanation that the head support is design in a way to not restrict head and neck movements. A service order was placed for hatch and buckle issue. The service consultant who responded to the service order, reported that additional in servicing was required by the customer and that there was nothing wrong with the bed. The bed is currently on rent at another account. We report this incident taking into account an allegation of buckle being stuck and due to potential for a death or serious injury if the malfunction was to reoccur.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007420694-2018-00089
MDR Report Key7424704
Report SourceUSER FACILITY
Date Received2018-04-12
Date of Report2018-05-11
Date of Event2018-03-16
Date Mfgr Received2018-04-27
Date Added to Maude2018-04-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KINGA STOLINSKA
Manufacturer StreetUL. KS. WAWRZYNIAKA 2
Manufacturer CityKOMORNIKI, 62-052, P
Manufacturer CountryPL
Manufacturer Postal62-052, PL
Manufacturer G1ARJOHUNTLEIGH, INC.
Manufacturer Street4958 STOUT DRIVE
Manufacturer CitySAN ANTONIO TX 78219
Manufacturer CountryUS
Manufacturer Postal Code78219
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameROTOPRONE
Generic NameBED, PATIENT ROTATION, POWERED
Product CodeIKZ
Date Received2018-04-12
Model Number209500
OperatorOTHER CAREGIVERS
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerARJOHUNTLEIGH, INC.
Manufacturer Address4958 STOUT DRIVE SAN ANTONIO TX 78219 US 78219


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-04-12