PANORAMIC X-RAY MODEL PC 1000 800724-3

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,other report with the FDA on 2018-04-12 for PANORAMIC X-RAY MODEL PC 1000 800724-3 manufactured by Young Innovations.

Event Text Entries

[105691053] The panoramic certified technician from tlm was doing a pm and found the proper set screws for the tube head to be missing and incorrect set screws were in the tube head. The tube head did not fall again between service calls. The correct set screws were ordered and the technician went back to the office to install them and align the tube head. Pm done. Tube head is correctly secured and realigned. No leaks found in the tube head. Resolved. Closed ad. Quality determination: the fall of the tube head could have potentially caused or contributed serious injury. Therefore, the incident has been deemed an mdr reportable event to fda.
Patient Sequence No: 1, Text Type: N, H10


[105691054] Dr. (b)(6) told panoramic that they were re-aligning the tube head between pan and ceph and the tube head fell off the arm. It did not hit or injure any staff/patient members. The office technician mounted the tube head back onto the machine and now appear to have alignment issues.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1832462-2018-00002
MDR Report Key7426087
Report SourceHEALTH PROFESSIONAL,OTHER
Date Received2018-04-12
Date of Report2018-04-12
Date of Event2018-03-26
Date Mfgr Received2018-03-26
Device Manufacturer Date2003-03-20
Date Added to Maude2018-04-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactADRIENNE DEMLAND
Manufacturer Street4321 GOSHEN RD.
Manufacturer CityFORT WAYNE IN 46818
Manufacturer CountryUS
Manufacturer Postal46818
Manufacturer Phone8006542027
Manufacturer G1YOUNG INNOVATIONS
Manufacturer Street4321 GOSHEN RD.
Manufacturer CityFORT WAYNE IN 46818
Manufacturer CountryUS
Manufacturer Postal Code46818
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePANORAMIC X-RAY MODEL PC 1000
Generic NamePC 1000
Product CodeEHD
Date Received2018-04-12
Model Number800724-3
OperatorDENTIST
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerYOUNG INNOVATIONS
Manufacturer Address4321 GOSHEN RD. FORT WAYNE IN 46818 US 46818


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-04-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.