DCI EQUIPMENT DC1335

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-04-12 for DCI EQUIPMENT DC1335 manufactured by Dci Equipment.

Event Text Entries

[105536465] The chair pcb is located in a fire resistant housing per the required electrical and safety requirements of iec 60601-1 (medical electrical equipment - part 1: general requirements for basic safety and essential performance). Dci equipment contacted the dental facility numerous times trying to obtain more information regarding this event as well as to request the part back for evaluation however dci equipment was continually routed back to a call center for the dental practice on all attempts made. Several messages were left with the facility but they never returned our calls. Therefore we are unable to evaluate the failed component.
Patient Sequence No: 1, Text Type: N, H10


[105536466] It was reported by the doctor's service technician that a dci equipment dental chair's circuit board located in the chair base assembly started to smoke and caught fire. There were no injuries reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1017522-2018-00010
MDR Report Key7427016
Date Received2018-04-12
Date of Report2018-03-15
Date Mfgr Received2018-03-15
Device Manufacturer Date2014-12-01
Date Added to Maude2018-04-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationSERVICE PERSONNEL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. FRANK RAY
Manufacturer Street11727 FRUEHAUF DRIVE
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal28273
Manufacturer Phone7045877227
Manufacturer G1KAVO DENTAL TECHNOLOGIES, LLC
Manufacturer Street11727 FRUEHAUF DRIVE
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal Code28273
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameDCI EQUIPMENT
Generic NameDENTAL CHAIR
Product CodeNRU
Date Received2018-04-12
Model NumberDC1335
Catalog NumberDC1335
OperatorDENTIST
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerDCI EQUIPMENT
Manufacturer Address11727 FRUEHAUF DRIVE CHARLOTTE NC 28273 US 28273


Patients

Patient NumberTreatmentOutcomeDate
10 2018-04-12

© 2024 FDA.report