MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2018-04-13 for LIGHT FANTASTIC II HBT LFT II manufactured by Pelton & Crane.
[105372404]
The light is being disposed and is not being returned. The dental light is over 23 years old and past the life of the device.
Patient Sequence No: 1, Text Type: N, H10
[105372405]
It was reported that a pelton & crane dental light arm assembly broke causing the light arm to drop down and hit the patient. The dentist evaluated the patient and did not see any injuries however as a precaution the dentist sent the patient to the er for evaluation.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1017522-2018-00012 |
MDR Report Key | 7428263 |
Report Source | USER FACILITY |
Date Received | 2018-04-13 |
Date of Report | 2018-04-03 |
Date of Event | 2018-03-28 |
Date Mfgr Received | 2018-04-03 |
Device Manufacturer Date | 1994-09-01 |
Date Added to Maude | 2018-04-13 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | BIOMEDICAL ENGINEER |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MR. FRANK RAY |
Manufacturer Street | 11727 FRUEHAUF DRIVE |
Manufacturer City | CHARLOTTE NC 28273 |
Manufacturer Country | US |
Manufacturer Postal | 28273 |
Manufacturer G1 | KAVO DENTAL TECHNOLOGIES, LLC |
Manufacturer Street | 11727 FRUEHAUF DRIVE |
Manufacturer City | CHARLOTTE NC 28273 |
Manufacturer Country | US |
Manufacturer Postal Code | 28273 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | LIGHT FANTASTIC II |
Generic Name | DENTAL LIGHT |
Product Code | EAZ |
Date Received | 2018-04-13 |
Model Number | HBT |
Catalog Number | LFT II |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | PELTON & CRANE |
Manufacturer Address | 11727 FRUEHAUF DRIVE CHARLOTTE NC 28273 US 28273 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2018-04-13 |