Event Text Entries

[105773619] Investigation including root cause analysis is in progress. A supplemental mdr will be filed as necessary in accordance with 21 cfr 803. 56 when additional reportable information becomes available. The manufacturer internal reference number is: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[105773620] A customer reported observing a loose caster on the microscope before surgery.
Patient Sequence No: 1, Text Type: D, B5


[118616920] The system was examined and the reported event was confirmed. The caster was tightened to resolve the issue. The system was tested and found to meet product specifications. The microscope was manufactured on october 21, 2014. Based on qa assessment, the product met specifications at the time of release. The facility is a mobile surgical equipment service, and per information provided, the microscopes are moved frequently as it is transported between customer sites. Therefore, the root cause of the caster coming loose is likely attributed to customer transport/handling. However, with the information provided, this is unable to be determined conclusively. The root cause cannot be determined conclusively. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2028159-2018-00764
MDR Report Key7433684
Date Received2018-04-16
Date of Report2018-08-27
Date Mfgr Received2018-08-02
Device Manufacturer Date2014-10-21
Date Added to Maude2018-04-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. BRYAN BLAKE
Manufacturer Street6201 SOUTH FREEWAY MAIL STOP AB2-6
Manufacturer CityFORT WORTH TX 76134
Manufacturer CountryUS
Manufacturer Postal76134
Manufacturer Phone8176152230
Manufacturer G1ENDURE MEDICAL, INC.
Manufacturer Street1455 VENTURA DR
Manufacturer CityCUMMING GA 30040
Manufacturer CountryUS
Manufacturer Postal Code30040
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report0

Device Details

Brand NameLUXOR OPHTHALMIC MICROSCOPES
Generic NameMICROSCOPE, OPERATING & ACC., AC-POWERED, OPHTHALMIC
Product CodeHRM
Date Received2018-04-16
Model NumberE-71
Catalog Number8065752239
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerENDURE MEDICAL, INC.
Manufacturer Address1455 VENTURA DR CUMMING GA 30040 US 30040


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-04-16