COVERED CP STENT 427 CVRDCP8Z45

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-04-18 for COVERED CP STENT 427 CVRDCP8Z45 manufactured by Numed, Inc..

Event Text Entries

[106043862] The covered stent was returned and the stent is clean. Two adhesive spots are detached, otherwise the covering is undamaged. Four adhesive spots are visible on the covering. The stent has been crimped down. The zigs are twisted at the end of the stent where the detachment occurred. The zigs at this end of the stent are overlapped. The outer layer of glue for each of the detached adhesive spots is intact and attached to the covering. The detachment is confirmed. The zigs at the end of the stent were twisted and overlapped, which damaged the inner layer of glue for each of the detached adhesive spots, allowing the detachment to occur. This damage was caused by user handling during the crimping process. A sample from each lot of stents is tested for covering strength. The sample met the required testing strength for release of the lot.
Patient Sequence No: 1, Text Type: N, H10


[106043863] Covering became dislodged during mounting process.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1318694-2018-00004
MDR Report Key7439365
Date Received2018-04-18
Date of Report2018-04-18
Date of Event2018-03-21
Date Mfgr Received2018-03-27
Device Manufacturer Date2017-04-25
Date Added to Maude2018-04-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. NICHELLE LAFLESH
Manufacturer Street2880 MAIN STREET
Manufacturer CityHOPKINTON NY 12965
Manufacturer CountryUS
Manufacturer Postal12965
Manufacturer Phone3153284491
Manufacturer G1NUMED, INC.
Manufacturer Street2880 MAIN STREET
Manufacturer CityHOPKINTON NY 12965
Manufacturer CountryUS
Manufacturer Postal Code12965
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCOVERED CP STENT
Generic NameAORTIC STENT
Product CodePNF
Date Received2018-04-18
Returned To Mfg2018-04-17
Model Number427
Catalog NumberCVRDCP8Z45
Lot NumberCCP-0780
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerNUMED, INC.
Manufacturer Address2880 MAIN STREET HOPKINTON NY 12965 US 12965


Patients

Patient NumberTreatmentOutcomeDate
10 2018-04-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.