TRIEEA28MT

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-04-18 for TRIEEA28MT manufactured by Us Surgical Puerto Rico.

Event Text Entries

[105875090] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[105875091] According to the reporter: the patient underwent an laparoscopic sigmoidectomy procedure. The circular stapler was used during the last step of the surgery to create the anastomosis between the distal colon and rectum stump. There was poor staple formation. The handle was fully squeezed, there was no excessive tissue incorporated in the barrel, the instrument was not improperly matched, and the donuts were complete. The anvil could be tilted after firing, the anvil was not bent, and the stapler sizers were used. A leak test was performed using a coloscope and inter-operatively the staple line looked nice. An anastomotic leak occurred three days post operatively and required re-intervention. There was temporary injury as the leak needed to be fixed. There was tissue damage as the leak was over sewn. The event lead to or extended the patients hospitalization time by two days. There was no reported patient injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2647580-2018-01909
MDR Report Key7441385
Date Received2018-04-18
Date of Report2018-04-18
Date of Event2018-03-11
Date Mfgr Received2018-04-03
Date Added to Maude2018-04-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactLISA HERNANDEZ
Manufacturer Street60 MIDDLETOWN AVE.
Manufacturer CityNORTH HAVEN CT 06473
Manufacturer CountryUS
Manufacturer Postal06473
Manufacturer Phone2034925563
Manufacturer G1US SURGICAL PUERTO RICO
Manufacturer Street201 SABANETAS INDUSTRIAL PARK
Manufacturer CityPONCE PR 007164401
Manufacturer CountryUS
Manufacturer Postal Code007164401
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Generic NamePLATE, BASE, SHELLAC
Product CodeEEA
Date Received2018-04-18
Model NumberTRIEEA28MT
Catalog NumberTRIEEA28MT
Lot NumberV7M0963X
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerUS SURGICAL PUERTO RICO
Manufacturer Address201 SABANETAS INDUSTRIAL PARK PONCE PR 007164401 US 007164401


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Required No Informationntervention 2018-04-18