COUCHMOUNT VARIAN EXACT * 40100-0A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2005-06-22 for COUCHMOUNT VARIAN EXACT * 40100-0A manufactured by Brainlab Ag.

Event Text Entries

[15038314] Medical center reported that their brainlab couchmount attached to the varian couch unintentionally detached from the couchmount and fell on the floor when the patient tried to sit up.
Patient Sequence No: 1, Text Type: D, B5


[15235368] Problem: the secondary locking mechanism on the couchmount is under certain conditions not reliable. Root cause: protruding magnets of the couch caused the secondary locking mechanism of the brainlab couchmount not to engage reliably. Corrective action: the magnets on the couch at the concerned medical center have been adjusted and the brainlab couchmount now sits flush to the couch and accordingly works reliably. Preventive actions: all couchmount customers will be informed to verify that their couchmount is locked reliably to the couch. The corresponding brainlab user manual will be updated to clearly describe the importance of this check. The design of the couchmount will be improved to clearly indicate when the secondary locking mechanism is engaged/reliable or not.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8043933-2005-00002
MDR Report Key744439
Report Source05,06
Date Received2005-06-22
Date of Report2005-05-23
Date Mfgr Received2005-05-23
Device Manufacturer Date2005-04-01
Date Added to Maude2006-08-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactPER PERSSON
Manufacturer StreetAMMERTHALSTRASSE 8
Manufacturer CityHEIMSTETTEN 85551
Manufacturer CountryGM
Manufacturer Postal85551
Manufacturer Phone899915680
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Remedial ActionNO
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOUCHMOUNT VARIAN EXACT
Generic NameCOUCHMOUNT FOR RADIOTHERAPY COUCH
Product CodeJAI
Date Received2005-06-22
Model Number*
Catalog Number40100-0A
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key732246
ManufacturerBRAINLAB AG
Manufacturer AddressAMMERTHALSTRASSE 8 HEIMSTETTEN GM 85551
Baseline Brand NameCOUCHMOUNT VARIAN EXACT
Baseline Generic NameCOUCHMOUNT FOR RADIOTHERAPY COUCH
Baseline Model No*
Baseline Catalog No40100-0A
Baseline ID*


Patients

Patient NumberTreatmentOutcomeDate
10 2005-06-22

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.