VANCOMYCIN 06779336190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-04-20 for VANCOMYCIN 06779336190 manufactured by Roche Diagnostics.

Event Text Entries

[106270769] (b)(4). (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[106270770] The customer complained of questionable low vanc3 vancomycin (vanc3) results for 1 patient tested on a cobas 8000 c 502 module compared to abbott architect. The patient had 4 separate sample that were tested on both the cobas c502 and the abbott architect. For sample 1 the cobas c502 result was < 4 mg/l. The abbott architect result for sample 1 was about 20 mg/l. For sample 2 on (b)(6) 2018 the cobas c502 result was < 4 mg/l. The abbott architect result for sample 2 was about 20 mg/l. For sample 3 on (b)(6) 2018 the cobas c502 result was < 4 mg/l. The abbott architect result for sample 3 was about 20 mg/l. For sample 4 on (b)(6) 2018 the cobas c502 result was 39 mg/l. The abbott architect result for sample 4 was about 80 mg/l. The cobas c502 results were released outside of the laboratory, but were considered not believable as the patient was being administered vancomycin. There was no allegation of an adverse event. No treatment was administered based on the cobas c502 results. The cobas c502 serial number was (b)(4). Interference in the sample was suspected. The investigation is currently ongoing.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-01231
MDR Report Key7446250
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-04-20
Date of Report2018-07-11
Date of Event2018-03-31
Date Mfgr Received2018-04-09
Date Added to Maude2018-04-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameVANCOMYCIN
Generic NameRADIOIMMUNOASSAY, VANCOMYCIN
Product CodeLEH
Date Received2018-04-20
Model NumberNA
Catalog Number06779336190
Lot Number260684
ID NumberNA
Device Expiration Date2019-01-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-04-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.