VITEK? 2 GRAM-POSITIVE (GP) IDENTIFICATION (ID) TEST KIT 21342

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-04-20 for VITEK? 2 GRAM-POSITIVE (GP) IDENTIFICATION (ID) TEST KIT 21342 manufactured by Biom?rieux, Inc..

Event Text Entries

[106861972] A customer in the united states contacted biom? Rieux to report the occurrence of misidentification of streptococcus constellatus (cap survey strain2017 d-17) as streptococcus anginosus in association with the vitek? 2 gp id test kit. Upon notification from cap, the customer repeated the test; obtained a low discrimination identification strep. Anginosus / strep. Constellatus ssp constellatus / strep. Gordonii. There is no indication or report from the laboratory that the discrepant result led to any adverse event related to any patient's state of health. There was no patient directly associated with the cap survey strain. Biom? Rieux investigation has been initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1950204-2018-00154
MDR Report Key7448263
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-04-20
Date of Report2018-05-15
Date Mfgr Received2018-04-24
Device Manufacturer Date2017-09-27
Date Added to Maude2018-04-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. ELLEN WELTMER
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147317301
Manufacturer G1BIOM
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal Code63042
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVITEK? 2 GRAM-POSITIVE (GP) IDENTIFICATION (ID) TEST KIT
Generic NameVITEK? 2 GP ID CARD
Product CodeLQL
Date Received2018-04-20
Catalog Number21342
Lot Number2420488103
Device Expiration Date2019-03-29
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOM?RIEUX, INC.
Manufacturer Address595 ANGLUM ROAD HAZELWOOD MO 63042 US 63042


Patients

Patient NumberTreatmentOutcomeDate
10 2018-04-20

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.