OAE PEDIATRIC FOAM TIP 203211

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2018-04-20 for OAE PEDIATRIC FOAM TIP 203211 manufactured by Natus Medical Incorporated.

Event Text Entries

[106148338] The complainant was provided replacement oae pediatric foam tips. Instructions for use provided with hearing devices compatible with oae pediatric foam tips state to not pinch the foam while attaching the ear tip onto the nozzle. For removing the foam ear tips, instructions state to not pull on the foam when removing the foam ear tips and caution that users may end up pulling the foam off the clear tube while the tube remains on the nozzle. Users are also instructed to remove, discard and replace probe tips before each test. Natus is investigating this issue through corrective actions at this time.
Patient Sequence No: 1, Text Type: N, H10


[106148339] Natus received a report that foam separated from the plastic tubing. The report concerned one bag of oae pediatric foam tips. The lot number was requested by natus technical service but the complainant could not provide it. There was no report of death, serious injury, delay in treatment, or environmental/safety concerns.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3018859-2018-00116
MDR Report Key7449253
Report SourceDISTRIBUTOR
Date Received2018-04-20
Date of Report2018-03-22
Date Mfgr Received2018-08-01
Date Added to Maude2018-04-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAMES FITZGERALD
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone2062685173
Manufacturer G1NATUS MEDICAL INCORPORATED
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal Code98108
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOAE PEDIATRIC FOAM TIP
Generic NameOAE PEDIATRIC FOAM TIP
Product CodeGWJ
Date Received2018-04-20
Model Number203211
Device Availability*
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INCORPORATED
Manufacturer Address5900 FIRST AVENUE SOUTH SEATTLE WA 98108 US 98108


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-04-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.