B5LT

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-04-23 for B5LT manufactured by Medline Renewal.

Event Text Entries

[106194851] It was reported that the clear plastic tip of a reprocessed ethicon b5lt 5mm trocar broke off as the surgeon was pulling the trocar out of the cannula, which was still in the patient's abdomen. The surgeon was able to locate the detached tip and retrieved it from the patient without further incident. There was no report of adverse patient consequence or prolonged anesthesia related to the incident. The sample was returned for evaluation and it was confirmed that the device was reprocessed. The complaint was confirmed however; a root cause could not be determined at this time. Due to the reported incident and in abundance of caution, this medwatch is being filed. If additional relevant information becomes available, a supplemental medwatch will be filed.
Patient Sequence No: 1, Text Type: N, H10


[106194852] It was reported that the clear plastic tip of a reprocessed ethicon b5lt 5mm trocar broke.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3032391-2018-00003
MDR Report Key7450333
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-04-23
Date of Report2018-04-23
Date of Event2018-03-29
Date Mfgr Received2018-03-30
Device Manufacturer Date2017-12-19
Date Added to Maude2018-04-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEPHEN WILSON
Manufacturer Street1500 NE HEMLOCK AVE
Manufacturer CityREDMOND OR 97756
Manufacturer CountryUS
Manufacturer Postal97756
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Generic NameETHICON ENDOPATH? XCEL? BLADELESS TROCAR, W/STABILITY SLEEVE, 100MM LENGTH 5.0MM
Product CodeNLM
Date Received2018-04-23
Returned To Mfg2018-04-13
Catalog NumberB5LT
Lot Number369119
OperatorPHYSICIAN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDLINE RENEWAL
Manufacturer Address1500 NE HEMLOCK AVE REDMOND OR 97756 US 97756


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2018-04-23

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.