NOMAD PRO 2 0.850.0022

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2018-04-23 for NOMAD PRO 2 0.850.0022 manufactured by Aribex.

Event Text Entries

[106258912] The unit was returned and an evaluation was conducted. Upon visual inspection, it appears that a thermal event occurred. It is not possible to determine the exact sequence of events that led to the thermal event. However, there is evidence that a short occured in the upper battery pack between the cells and printed circuit board. There was a blackening on the top of the upper battery pack and its printed circuit board (pcb). Cells 5 and 6 appear to have been involved in the thermal event, likely by supplying energy into a fault in the board. Additionally, there was a slight melting of the handset plastic enclosure. This concludes the investigation.
Patient Sequence No: 1, Text Type: N, H10


[106258913] It was reported that the device and battery were smoking while not in use. There was no report of injuries, patient or user involvement and no impact to patient care.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1017522-2018-00015
MDR Report Key7452805
Report SourceUSER FACILITY
Date Received2018-04-23
Date of Report2018-04-23
Date of Event2018-03-29
Date Mfgr Received2018-03-29
Device Manufacturer Date2014-03-03
Date Added to Maude2018-04-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. ERIKA MARTIN
Manufacturer Street11727 FRUEHAUF DRIVE
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal28273
Manufacturer G1ARIBEX
Manufacturer Street11727 FRUEHAUF DRIVE
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal Code28273
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNOMAD PRO 2
Generic NameEXTRAORAL SOURCE X-RAY SYSTEM
Product CodeEHD
Date Received2018-04-23
Returned To Mfg2018-04-11
Model Number0.850.0022
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerARIBEX
Manufacturer Address11727 FRUEHAUF DRIVE CHARLOTTE NC 28273 US 28273


Patients

Patient NumberTreatmentOutcomeDate
10 2018-04-23

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.