MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-04-24 for TRIGL TRIGLYCERIDES 05171407190 manufactured by Roche Diagnostics.
[107092526]
Unique identifier (udi)# (b)(4). The event occurred in: (b)(6).
Patient Sequence No: 1, Text Type: N, H10
[107092527]
The customer complained of a questionable result for 1 patient tested for trigl triglycerides (trigl) on a cobas 8000 c (701) module. The initial trigl result was 182 mg/dl. The same sample was retested with a 1:20 dilution with a result of 10683 mg/dl. The erroneous result was released outside of the laboratory. There was no allegation of an adverse event. The trigl reagent lot was 25141601 with an expiration date of 30-jun-2018. The investigation is currently ongoing.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1823260-2018-01266 |
MDR Report Key | 7454646 |
Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
Date Received | 2018-04-24 |
Date of Report | 2018-05-10 |
Date of Event | 2018-04-11 |
Date Mfgr Received | 2018-04-12 |
Date Added to Maude | 2018-04-24 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | NA MICHAEL LESLIE |
Manufacturer Street | 9115 HAGUE ROAD NA |
Manufacturer City | INDIANAPOLIS IN 46250 |
Manufacturer Country | US |
Manufacturer Postal | 46250 |
Manufacturer Phone | 3175214343 |
Manufacturer G1 | ROCHE DIAGNOSTICS GMBH |
Manufacturer Street | SANDHOFERSTRASSE 116 NA |
Manufacturer City | MANNHEIM (BADEN-WURTTEMBERG) 68305 |
Manufacturer Country | GM |
Manufacturer Postal Code | 68305 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 3 |
Brand Name | TRIGL TRIGLYCERIDES |
Generic Name | TRIGLYCERIDES GPO WITHOUT FREE GLYCEROL |
Product Code | CDT |
Date Received | 2018-04-24 |
Model Number | NA |
Catalog Number | 05171407190 |
Lot Number | 25141601 |
ID Number | NA |
Device Expiration Date | 2018-06-30 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2018-04-24 |