MAUDE MDR 7454646

MDR report key
7454646
Report number
1823260-2018-01266
Event key
0
Event type
3
Date of event
2018-04-11
Date received
2018-04-24
Adverse event
3
Product problem
3
Patients in event
0
Reporter occupation
0
Health professional
3
Initial report to FDA
3
Event location
3

Manufacturer Contact#

Contact
NA MICHAEL LESLIE
Address
9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US
Phone
317-317-3175
Report source
M
Manufacturer link flag
Y

Devices#

Seq, Brand, Generic table
SeqBrandGenericManufacturerProduct codeModelCatalogLotPMA510(k)ImplantEvaluatedAvailability
1TRIGL TRIGLYCERIDESTRIGLYCERIDES GPO WITHOUT FREE GLYCEROLROCHE DIAGNOSTICSCDTNA0517140719025141601R Y

Patients#

Sequence, Received, Treatment table
SequenceReceivedTreatmentOutcome
12018-04-240

Event Narratives#

N

Patient 1

UNIQUE IDENTIFIER (UDI)# (B)(4). THE EVENT OCCURRED IN: (B)(6).

D

Patient 1

THE CUSTOMER COMPLAINED OF A QUESTIONABLE RESULT FOR 1 PATIENT TESTED FOR TRIGL TRIGLYCERIDES (TRIGL) ON A COBAS 8000 C (701) MODULE. THE INITIAL TRIGL RESULT WAS 182 MG/DL. THE SAME SAMPLE WAS RETESTED WITH A 1:20 DILUTION WITH A RESULT OF 10683 MG/DL. THE ERRONEOUS RESULT WAS RELEASED OUTSIDE OF THE LABORATORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE TRIGL REAGENT LOT WAS 25141601 WITH AN EXPIRATION DATE OF 30-JUN-2018. THE INVESTIGATION IS CURRENTLY ONGOING.