HYDRA SIT-355-15

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2018-04-26 for HYDRA SIT-355-15 manufactured by Boston Scientific - Marlborough.

Event Text Entries

[106707558] Patient? S exact age is unknown; however, it was reported that the patient was over the age of 18. The reported lot number could not be matched to the reported device. Therefore, the lot expiration and device manufacture dates are unknown. However, the complainant reported that the device was not expired. (b)(4). The complainant indicated that the device was disposed and will not be returned for evaluation; therefore a failure analysis of the complaint device could not be completed. If any further relevant information is identified, a supplemental medwatch will be filed.
Patient Sequence No: 1, Text Type: N, H10


[106707559] It was reported to boston scientific corporation that a hydra irrigation tubing was used during a colonoscopy procedure performed on an unknown date. According to the complainant, during the procedure, the irrigation tubing was connected to the biopsy channel of the colonoscope with a needle-like connector. Upon irrigation, it was noted that there was backflow of stool through the backflow valve into the tubing and sterile water bottle. The tubing and water bottle were replaced and the procedure was completed. There were no patient complications reported as a result of this event. The patient's condition at the conclusion of the procedure was reported to be fine.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2018-01278
MDR Report Key7465140
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2018-04-26
Date of Report2018-04-05
Date Mfgr Received2018-04-05
Date Added to Maude2018-04-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEMP. NANCY CUTINO
Manufacturer Street100 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834000
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHYDRA
Generic NamePUMP, AIR, NON-MANUAL, FOR ENDOSCOPE
Product CodeFEQ
Date Received2018-04-26
Model NumberSIT-355-15
Lot Number170630
OperatorPHYSICIAN
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC - MARLBOROUGH
Manufacturer Address100 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
10 2018-04-26