COVE 53WH981U

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-04-27 for COVE 53WH981U manufactured by Wieland Healthcare.

Event Text Entries

[106750065]
Patient Sequence No: 1, Text Type: N, H10


[106750066] With orthopedic bariatric chair in locked position, upon patient pushing up to stand with left hand on armrest of chair, left side of chair began to slide. When weight is applied to one of the arms of the chair whether it is to stand up or sit down the chair has been known to slide away from the individual and could cause a patient to fall. The single locking mechanism does not lock all four wheels at the corners and it pivots on 2 wheels underneath the chair. Maintenance manager tried to get the chair to slide and was successful in his attempt. Two other members of the facilities maintenance team also caused it to slide as well. The manufacturer is wieland healthcare. Manufacturer response for reclining bariatric chair, recliner chair (per site reporter). Contacted (b)(6) at (b)(6). Setting up a meeting.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number7468224
MDR Report Key7468224
Date Received2018-04-27
Date of Report2018-03-28
Date of Event2018-03-21
Report Date2018-03-28
Date Reported to FDA2018-03-28
Date Reported to Mfgr2018-03-28
Date Added to Maude2018-04-27
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOVE
Generic NameCHAIR, WITH CASTERS
Product CodeINM
Date Received2018-04-27
Model Number53WH981U
OperatorNURSE
Device AvailabilityY
Device Age2 YR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerWIELAND HEALTHCARE
Manufacturer Address10785 ROSE AVE NEW HAVEN IN 46774 US 46774


Patients

Patient NumberTreatmentOutcomeDate
10 2018-04-27