3343N 3M? CAVILON? NO STING BARRIER FILM N/A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,consum report with the FDA on 2018-04-27 for 3343N 3M? CAVILON? NO STING BARRIER FILM N/A manufactured by 3m Health Care.

Event Text Entries

[106829626] The product is applied as a liquid and flammable until dry. The label includes a flammability warning and also instructs readers to the instructions for use which states the following warnings: danger! Highly flammable! Cavilon no sting barrier film is highly flammable until it has completely dried on the skin. Cavilon no sting barrier film should only be applied when no ignition sources or heat producing devices are in use. Avoid using around flames. Use in a well ventilated area. The hospital could not confirm a specific ignition source, 3m is still investigating the potential root cause for this reported event.
Patient Sequence No: 1, Text Type: N, H10


[106829627] A hospital employee reported 3343n 3m? Cavilon? No sting barrier film was applied to existing skin irritation at a patient's iv site on their arm. The skin allegedly ignited where the product was applied. The fire was quickly extinguished by the staff. The patient allegedly experienced a burn. The hospital could not confirm a specific ignition source.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2110898-2018-00044
MDR Report Key7470778
Report SourceCOMPANY REPRESENTATIVE,CONSUM
Date Received2018-04-27
Date of Report2018-04-27
Date of Event2018-04-01
Date Mfgr Received2018-04-03
Device Manufacturer Date2017-07-01
Date Added to Maude2018-04-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactBRENDAN CASEY, PH.D.
Manufacturer Street3M CENTER BUILDING 275-5W-06 2510 CONWAY AVE
Manufacturer CityST. PAUL MN 55144
Manufacturer CountryUS
Manufacturer Postal55144
Manufacturer Phone6517374488
Manufacturer G13M HEALTH CARE LIMITED
Manufacturer StreetDERBY ROAD
Manufacturer CityLOUGHBOROUGH, LEICESTERSHIRE LE115SF
Manufacturer CountryUK
Manufacturer Postal CodeLE11 5SF
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name3343N 3M? CAVILON? NO STING BARRIER FILM
Generic NameCAVILON NO STING BARRIER FILM
Product CodeKMF
Date Received2018-04-27
Model NumberN/A
Catalog Number3343N
Lot Number2020-07 KA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
Manufacturer3M HEALTH CARE
Manufacturer Address2510 CONWAY AVE ST. PAUL MN 55144 US 55144


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-04-27

© 2026 FDA.report