CLINICAL CHEMISTRY LACTATE DEHYDROGENASE 02P56-21

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2018-04-30 for CLINICAL CHEMISTRY LACTATE DEHYDROGENASE 02P56-21 manufactured by Abbott Manufacturing Inc.

Event Text Entries

[107599801] An evaluation is in process. A follow-up report will be submitted when the evaluation is complete. There is no further patient information provided due to privacy issues, other than the patient is receiving medical care and being monitored by health care professionals. Additional information for section a. Patient information; 1. Patient identifier- multiple = (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[107599802] The customer reported falsely elevated ldh results on four patients. The results provided were: on (b)(6) 2018 - sid(b)(6), initial ldh = 1597. 0 u/l (normal range 0-248u/l) / second sample =200 u/l; sid(b)(6) =298 / 207 / 195u/l; sid(b)(6) = 376 / 546 / 336u/l; sid(b)(6) = 329 / 219 / 224u/l. There was no reported impact to patient management.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1628664-2018-00167
MDR Report Key7472279
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2018-04-30
Date of Report2018-06-12
Date of Event2018-04-10
Date Mfgr Received2018-06-08
Device Manufacturer Date2018-01-22
Date Added to Maude2018-04-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT MANUFACTURING INC
Manufacturer Street1921 HURD DRIVE
Manufacturer CityIRVING TX 75038
Manufacturer CountryUS
Manufacturer Postal Code75038
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCLINICAL CHEMISTRY LACTATE DEHYDROGENASE
Generic NameLACTATE DEHYDROGENASE
Product CodeCFJ
Date Received2018-04-30
Catalog Number02P56-21
Lot Number75682UN17
Device Expiration Date2018-11-24
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MANUFACTURING INC
Manufacturer Address1921 HURD DRIVE IRVING TX 75038 US 75038


Patients

Patient NumberTreatmentOutcomeDate
10 2018-04-30