OPTUNE TFH-9100 N/A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer,health professional report with the FDA on 2018-04-30 for OPTUNE TFH-9100 N/A manufactured by Novocure, Ltd..

Event Text Entries

[106929837] Novocure medical opinion is that a contribution of device use to the event cannot be ruled out. Other contributing factors in this patient include: concomitant temozolomide (headache is listed as a common adverse event. Source: temozolomide prescribing information) and underlying disease (tumor progression reported in month prior to event). Headaches were reported on the ef-14 trial of optune together with temozolomide (tmz) compared to tmz alone in patients with newly diagnosed gbm in both arms of the trial (28% and 20% in optune/tmz and tmz arms respectively). Headache is a common symptom of disease in gbm.
Patient Sequence No: 1, Text Type: N, H10


[106929838] A (b)(6) male patient with newly diagnosed glioblastoma began optune therapy on (b)(6) 2017. On (b)(6) 2018, the patient reported that he had been hospitalized for one night due to headache experienced while using optune therapy. No further information was provided. The prescribing physician was contacted for additional information with no response.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3009453079-2018-00100
MDR Report Key7473924
Report SourceCONSUMER,HEALTH PROFESSIONAL
Date Received2018-04-30
Date of Report2018-06-05
Date of Event2018-04-09
Date Mfgr Received2018-05-16
Device Manufacturer Date2017-07-13
Date Added to Maude2018-04-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDR. EILON KIRSON
Manufacturer StreetTOPAZ BUILDING, SHA'AR HACARMEL 4TH FLOOR
Manufacturer CityHAIFA 31905
Manufacturer CountryUS
Manufacturer Postal31905
Manufacturer Phone9724850120
Manufacturer G1NOVOCURE, LTD.
Manufacturer StreetTOPAZ BUILDING, SHA'AR HACARMEL 4TH FLOOR
Manufacturer CityHAIFA, 31905
Manufacturer CountryIS
Manufacturer Postal Code31905
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOPTUNE
Generic NameOPTUNE
Product CodeNZK
Date Received2018-04-30
Model NumberTFH-9100
Catalog NumberN/A
Lot NumberN/A
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Age9 MO
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerNOVOCURE, LTD.
Manufacturer AddressTOPAZ BUILDING, SHA'AR HACARMEL 4TH FLOOR HAIFA, 31905 IS 31905


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2018-04-30