MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-04-30 for JAUNDICE METER JM105 N/A manufactured by Draeger Medical Systems, Inc..
[106954052]
There is no indication that the meter did not function as intended. The users in this case misinterpreted the displayed "-o-" display for an out of range bilirubin reading. As a result there was a delay in treatment for the patient (phototherapy, and double exchange transfusion). Draeger has initiated a field action.
Patient Sequence No: 1, Text Type: N, H10
[106954053]
Due to a processing error, this mdr is being submitted late. The customer reported that a patient suspected of being jaundiced was screened with the meter which displayed a blinking -0- and the patient was discharged due to the user misinterpreting the out of range bilirubin reading ("-o-") the jaundice meter displayed. There was a delay in treatment (phototherapy, and double exchange transfusion). It was reported the patient has been diagnosed with bilateral auditory neuropathy.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2510954-2018-00003 |
MDR Report Key | 7473939 |
Date Received | 2018-04-30 |
Date of Report | 2018-04-23 |
Date of Event | 2017-10-20 |
Date Mfgr Received | 2017-12-13 |
Device Manufacturer Date | 2015-08-25 |
Date Added to Maude | 2018-04-30 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MR. DHAVAL TRIVEDI |
Manufacturer Street | 3135 QUARRY ROAD |
Manufacturer City | TELFORD PA 18969 |
Manufacturer Country | US |
Manufacturer Postal | 18969 |
Manufacturer Phone | 9783798077 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | 1220063-04/17/2018-001-C |
Event Type | 3 |
Type of Report | 0 |
Brand Name | JAUNDICE METER |
Generic Name | NON-INVASIVE TRANSCUTANEOUS BILIRUBMINOMETER |
Product Code | MQM |
Date Received | 2018-04-30 |
Model Number | JM105 |
Catalog Number | N/A |
Lot Number | N/A |
Device Expiration Date | 2000-01-01 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | DRAEGER MEDICAL SYSTEMS, INC. |
Manufacturer Address | 3135 QUARRY ROAD TELFORD PA 18969 US 18969 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Deathisabilit | 2018-04-30 |