[106954052] There is no indication that the meter did not function as intended. The users in this case misinterpreted the displayed "-o-" display for an out of range bilirubin reading. As a result there was a delay in treatment for the patient (phototherapy, and double exchange transfusion). Draeger has initiated a field action.
Patient Sequence No: 1, Text Type: N, H10


[106954053] Due to a processing error, this mdr is being submitted late. The customer reported that a patient suspected of being jaundiced was screened with the meter which displayed a blinking -0- and the patient was discharged due to the user misinterpreting the out of range bilirubin reading ("-o-") the jaundice meter displayed. There was a delay in treatment (phototherapy, and double exchange transfusion). It was reported the patient has been diagnosed with bilateral auditory neuropathy.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2510954-2018-00003
MDR Report Key7473939
Date Received2018-04-30
Date of Report2018-04-23
Date of Event2017-10-20
Date Mfgr Received2017-12-13
Device Manufacturer Date2015-08-25
Date Added to Maude2018-04-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. DHAVAL TRIVEDI
Manufacturer Street3135 QUARRY ROAD
Manufacturer CityTELFORD PA 18969
Manufacturer CountryUS
Manufacturer Postal18969
Manufacturer Phone9783798077
Single Use3
Previous Use Code3
Removal Correction Number1220063-04/17/2018-001-C
Event Type3
Type of Report0

Device Details

Brand NameJAUNDICE METER
Generic NameNON-INVASIVE TRANSCUTANEOUS BILIRUBMINOMETER
Product CodeMQM
Date Received2018-04-30
Model NumberJM105
Catalog NumberN/A
Lot NumberN/A
Device Expiration Date2000-01-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerDRAEGER MEDICAL SYSTEMS, INC.
Manufacturer Address3135 QUARRY ROAD TELFORD PA 18969 US 18969


Patients

Patient NumberTreatmentOutcomeDate
101. Deathisabilit 2018-04-30

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.