[107573420] Numerous attempts were made to contact the customer to obtain the sample but a response was not received. The dhr was reviewed and it was confirmed that lot d172361 had (b)(4) cases that were manufactured from 08/24/17 to 08/26/17. No defects were reported during quality inspection. Based on the device history record this does not appear to be the result of a personnel, process or material issue. Since no sample was provided the non conformity could not be confirmed. No additional actions will be taken at this time. A follow up report will be submitted if additional information becomes available.
Patient Sequence No: 1, Text Type: N, H10


[107573421] During case, circulator noticed saline in warmer not contained in sterile drape. No patient injury or treatment was reported for the incident.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8043817-2018-00014
MDR Report Key7476239
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-05-01
Date of Report2018-04-13
Date of Event2018-02-14
Date Mfgr Received2018-04-13
Date Added to Maude2018-05-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS JENNIFER MRKVICKA
Manufacturer Street1 ECOLAB PLACE
Manufacturer CityST. PAUL MN 55102
Manufacturer CountryUS
Manufacturer Postal55102
Manufacturer Phone6512502237
Manufacturer G1MICROTEK DOMINICANA
Manufacturer StreetZONA FRANCA NO. 2
Manufacturer CityLA ROMANA,
Manufacturer CountryDR
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameORS-320 SLUSH WARMER DISC DRAPE
Generic NameFLUID WARMING AND SLUSH DRAPES
Product CodeLHC
Date Received2018-05-01
Model NumberORS-320
Lot NumberD172361
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMICROTEK DOMINICANA, S.A.
Manufacturer AddressZONA FRANCA NO. 2 LA ROMANA, DR


Patients

Patient NumberTreatmentOutcomeDate
10 2018-05-01

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.