STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL M0063903200 390-320

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2018-05-01 for STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL M0063903200 390-320 manufactured by Boston Scientific - Marlborough.

Event Text Entries

[107193167] Patient's exact age is unknown; however it was reported that the patient was over the age of 18. (b)(4). A stone cone nitinol retrieval coil was received for analysis. A visual analysis of the returned device, found that the blue sheath overrode the distal stop. As a result, the device is stuck closed and will not open. The blue/green heat shrink is pulled away from the distal stop. The condition of the returned device confirms the event of device won? T open. The sheath most likely overrode the distal stop during preparation and outside the patient. Therefore the most probable root cause is? Handling damage". The complaint was caused by handling of the device or portion of the device without direct patient contact either during unpacking, preparation, or shipping; at the end of the procedure; or when packaging for return. A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications at the time of release for distribution.
Patient Sequence No: 1, Text Type: N, H10


[107193168] It was reported to boston scientific corporation that a stone cone nitinol urological retrieval coil was used in the ureter during a ureteroscopy procedure performed on (b)(6) 2018. According to the complainant, during introduction, the coil failed to opened. The procedure was completed with another stone cone. There were no patient complications reported as a result of this event. The patient's condition at the conclusion of the procedure was reported to be? Stable?. This event has been deemed a reportable event based on the investigation results; stone cone retrieval coil coating peeled/shared/frayed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2018-01159
MDR Report Key7477963
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2018-05-01
Date of Report2018-02-28
Date of Event2018-02-27
Date Mfgr Received2018-04-05
Device Manufacturer Date2017-07-10
Date Added to Maude2018-05-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEMP. NANCY CUTINO
Manufacturer Street100 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834000
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTONE CONE NITINOL UROLOGICAL RETRIEVAL COIL
Generic NameDISLODGER, STONE, FLEXIBLE
Product CodeFGO
Date Received2018-05-01
Returned To Mfg2018-03-05
Model NumberM0063903200
Catalog Number390-320
Lot Number0003809705
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC - MARLBOROUGH
Manufacturer Address100 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
10 2018-05-01

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.