CLINICAL CHEMISTRY GLUCOSE 03L82-21

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2018-05-01 for CLINICAL CHEMISTRY GLUCOSE 03L82-21 manufactured by Abbott Manufacturing Inc.

Event Text Entries

[107716008] An evaluation is in process. A follow-up report will be submitted when the evaluation is complete.
Patient Sequence No: 1, Text Type: N, H10


[107716009] The customer observed multiple falsely depressed glucose results while using clinical chemistry glucose reagents. The following data was provided for the same patient: sid: (b)(6) (serum) initial 47 mg/dl, repeat 47 mg/dl. Glucometer reading 15 minutes earlier was higher, 158 mg/dl. Previous results for the patient using the clinical chemistry glucose method ranged from 44 to 47 mg/dl. The customer stated the patient takes vitamin d, a multivitamin, fenlasaxine (150 mg, 2 tablets per day) and n-acetyl-l-cysteine (600 mg, 4 tablets per day). No change to medication was reported due to the falsely depressed glucose results. No impact to patient management was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1628664-2018-00170
MDR Report Key7478071
Report SourceHEALTH PROFESSIONAL
Date Received2018-05-01
Date of Report2018-06-05
Date of Event2018-04-06
Date Mfgr Received2018-05-11
Device Manufacturer Date2017-09-01
Date Added to Maude2018-05-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT MANUFACTURING INC
Manufacturer Street1921 HURD DRIVE
Manufacturer CityIRVING TX 75038
Manufacturer CountryUS
Manufacturer Postal Code75038
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCLINICAL CHEMISTRY GLUCOSE
Generic NameGLUCOSE
Product CodeCFR
Date Received2018-05-01
Catalog Number03L82-21
Lot Number51494UQ07
Device Expiration Date2018-07-31
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MANUFACTURING INC
Manufacturer Address1921 HURD DRIVE IRVING TX 75038 US 75038


Patients

Patient NumberTreatmentOutcomeDate
10 2018-05-01