DISPOSABLE ECKARDT TWINLIGHT CHANDELIER. (27 GAUGE / 0.4MM) 3269.MBD27

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-05-02 for DISPOSABLE ECKARDT TWINLIGHT CHANDELIER. (27 GAUGE / 0.4MM) 3269.MBD27 manufactured by Dutch Opthalmic Research Center Bv.

Event Text Entries

[107473144] Investigation could not reveal the root cause, due to severe contamination. Since it is containing silicon dirt, foreign particles, can cause foreign body reaction to patient, hence it is concluded as reportable. This reportable event was identified during a voluntary retrospective review of all complaints since 2015 by the manufacturer (b)(4) details of this activity were discussed with cdrh office of compliance (b)(4) ) during a tele-conference on (b)(4) 2017. All available information has been provided. This complaint has now been closed.
Patient Sequence No: 1, Text Type: N, H10


[107473145] A report has been received of an event that occurred in (b)(6): there have been two particles which are probably from the silicon medical grade. It`s difficult to recognize at the photos but we could definitely see them. She tried to save them but they have been too thin so we lost them.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1222074-2018-00092
MDR Report Key7480345
Date Received2018-05-02
Date of Report2015-04-28
Date of Event2015-04-28
Date Mfgr Received2018-04-28
Date Added to Maude2018-05-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS LAURA SMITH
Manufacturer Street10 CONTINENTAL DR BLDG 1
Manufacturer CityEXETER NH 03833
Manufacturer CountryUS
Manufacturer Postal03833
Manufacturer Phone6037786929
Manufacturer G1DUTCH OPTHALMIC RESEARCH CENTER BV
Manufacturer StreetSCHEIJDELVEWEG 2
Manufacturer CityZUIDLAND, 3214VN
Manufacturer CountryNL
Manufacturer Postal Code3214VN
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameDISPOSABLE ECKARDT TWINLIGHT CHANDELIER. (27 GAUGE / 0.4MM)
Generic NameDISPOSABLE ECKARDT TWINLIGHT CHANDELIER. (27 GAUGE / 0.4MM)
Product CodeMPA
Date Received2018-05-02
Catalog Number3269.MBD27
Lot Number2000368813
OperatorPHYSICIAN
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerDUTCH OPTHALMIC RESEARCH CENTER BV
Manufacturer AddressSCHEIJDELVEWEG 2 ZUIDLAND, 3214VN NL 3214VN


Patients

Patient NumberTreatmentOutcomeDate
10 2018-05-02