AFINION 2

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-05-04 for AFINION 2 manufactured by Alere Technologies As.

Event Text Entries

[107693992] An error has been detected in the afinion 2 software used in combination with the alere afinion acr test. Afinion 2 is cleared in the usa but has not yet been placed on the market and the problem described has only occured in markets outside u. S. The software defect results in deactivation of the temperature correction of the instrument. This temperature correction for the albumine measurements is required for the alere afinion acr test. The potential hazard is erroneous results for albumin and acr. The lack of temperature correction might give false low results at analyzer temperatures below 27 deg c and false high results above 27 deg c. The deviation from correct result is dependant on the actual analyzer temperature and thereby the room temperature where the analyzer is plazed. The table below shows the estimated impact on test results due to lack of temperature correction. Afinion 2 analyzer: temperature: 19, 20, 21, 22, 23, 24, 25, 26, 27, 28, 29, 30, 31, 32, 33, 34, 35, 36, 37. Deviation of albumin and acr result due to sw error: -18 %, -16 %, -14 %, -12 %, -9 %, -7 %, -5 %, -2 %, 0 %, 2 %, 5 %, 7 %, 9 %, 12 %, 14 %, 16 %, 18 %, 21 %, 23 %.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9613069-2018-00001
MDR Report Key7486427
Date Received2018-05-04
Date Mfgr Received2018-02-27
Date Added to Maude2018-05-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactMS MONICA VALLESTAD
Manufacturer StreetKJELSAASVEIEN 161
Manufacturer CityOSLO, 0504
Manufacturer CountryNO
Manufacturer Postal0504
Manufacturer G1ALERE TECHNOLOGIES AS
Manufacturer StreetP.O.BOX 6863 RODEL
Manufacturer CityOSLO, 0504
Manufacturer CountryNO
Manufacturer Postal Code0504
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameAFINION 2
Generic NameANALYZER
Product CodeJQT
Date Received2018-05-04
OperatorMEDICAL ASSISTANT
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerALERE TECHNOLOGIES AS
Manufacturer AddressP.O.BOX 6863 RODEL?KKA OSLO, 0504 NO 0504


Patients

Patient NumberTreatmentOutcomeDate
10 2018-05-04