CONFIDENCE HYDRAULIC PUMP 283906100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-05-04 for CONFIDENCE HYDRAULIC PUMP 283906100 manufactured by Medos International Sàrl Ch.

Event Text Entries

[107999145] (b)(4). Visual examination of the hydraulic pump revealed that the pump's tube has broken into two pieces, with the handle mechanism attached to one of the main pieces. Also, the hose connecting the hydraulic cylinder and the rectus connector is separated. No difficulties engaging and disengaging the rectus connector and the quick-connect feature of the cement reservoir connection were detected. Visual observations identified that there was hardened cement inside the confidence injector body. A review of the device history record was conducted. No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer. All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting. The root cause of the pump breaking during use cannot be determined from the sample and the information provided. A potential root cause may be imperfections cause while staking the pump? S tube to the inside of the body of the pump. Functional analysis of the hydraulic pump revealed no difficulties engaging and disengaging the rectus connector and the quick-connect feature of the cement reservoir connection. Therefore, this investigation could not verify or identify any evidence of product contribution to the reported problem. As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
Patient Sequence No: 1, Text Type: N, H10


[107999146] It was reported that on (b)(6) 2018, the support staff had mixed the cement and they were getting ready to attach the cap to the vial and the metal connector would not latch on to the cap. Scrub personnel and surgeons all tried without success. It was deemed useless and a new complete box was opened. Cement only has a life of 9 minutes, so they could not use the cement due to this failure. All scrub personnel and surgeons used new clean over gloves when dealing with the cement. Surgical delay and patient outcome unknown. Concomitant devices reported: confidence spinal cement system (part # 283910000, lot # 163392, quantity # 1). This complaint involves one (1) device.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1526439-2018-50453
MDR Report Key7487160
Date Received2018-05-04
Date of Report2018-02-13
Date of Event2018-02-13
Date Mfgr Received2018-04-23
Device Manufacturer Date2017-11-28
Date Added to Maude2018-05-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJASON BUSCH
Manufacturer Street325 PARAMOUNT DR
Manufacturer CityRAYNHAM MA 02767
Manufacturer CountryUS
Manufacturer Postal02767
Manufacturer Phone5088808201
Manufacturer G1MEDOS INTERNATIONSL SARL
Manufacturer StreetCHEMIN BLANC 38
Manufacturer CityLE LOCLE
Manufacturer CountrySZ
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCONFIDENCE HYDRAULIC PUMP
Generic NameGUIDE, NEEDLE, SURGICAL
Product CodeGDF
Date Received2018-05-04
Returned To Mfg2018-04-25
Catalog Number283906100
Lot Number167493
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDOS INTERNATIONAL SàRL CH
Manufacturer AddressCHEMIN-BLANC 38 LE LOCLE 02400 SZ 02400


Patients

Patient NumberTreatmentOutcomeDate
10 2018-05-04