FORCEPS MCEN116 DIA 4.0MM VAUGHAN PUNCH

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-05-04 for FORCEPS MCEN116 DIA 4.0MM VAUGHAN PUNCH manufactured by Integra Microfrance S.a.s..

Event Text Entries

[107588187] The device was not yet returned to the manufacturer for analysis. The plant investigation is in progress and a supplemental medwatch report will be submitted upon completion of the investigation.
Patient Sequence No: 1, Text Type: N, H10


[107588188] A field sales manager reported on behalf of the customer that on (b)(6) 2017, the tip of the mcen116 forceps mcen116 dia 4. 0mm vaughan punch was broken. The product was placed in position at sphenoid face and a bite was attempted on the medical sphenoid face. The bone was fairly thin but the tip broke instantly and fell into the sphenoid sinus. However, no patient injury, no medical revision or intervention was done and no delay in surgery was reported. Additional information has been requested but no other clinical information has been provided.
Patient Sequence No: 1, Text Type: D, B5


[132629777] The device was not returned to the manufacturer for physical evaluation, therefore the failure mode cannot be confirmed. The serial/lot number was not provided therefore a dhr review and investigation could not be performed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2523190-2018-00068
MDR Report Key7487823
Date Received2018-05-04
Date of Report2018-04-04
Date of Event2018-04-03
Date Mfgr Received2018-05-16
Date Added to Maude2018-05-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactUSER KIMBERLY SHELLY
Manufacturer Street311 ENTERPRISE DRIVE
Manufacturer CityPLAINSBORO NJ 08536
Manufacturer CountryUS
Manufacturer Postal08536
Manufacturer Phone6099362393
Manufacturer G1INTEGRA MICROFRANCE S.A.S.
Manufacturer StreetLE PAVILLON
Manufacturer CitySAINT AUBIN LE MONIAL 03160
Manufacturer CountryFR
Manufacturer Postal Code03160
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameFORCEPS MCEN116 DIA 4.0MM VAUGHAN PUNCH
Generic NamePFM11
Product CodeKAY
Date Received2018-05-04
Catalog NumberMCEN116
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerINTEGRA MICROFRANCE S.A.S.
Manufacturer AddressLE PAVILLON LE PAVILLON SAINT AUBIN LE MONIAL 03160 FR 03160


Patients

Patient NumberTreatmentOutcomeDate
10 2018-05-04

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.