MEDLINE MDT2168210

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-05-07 for MEDLINE MDT2168210 manufactured by Medline Industries, Inc..

Event Text Entries

[107483848]
Patient Sequence No: 1, Text Type: N, H10


[107483849] Patient was prepped and draped to have a dialysis catheter installed on right upper neck. Sterile towels were draped around incisional area. While doctor was trying to advance dialysis catheter over wire he encountered difficulty and the catheter was unable to make a smooth transition. It was noted there was a string wrapped around the catheter preventing the catheter to be advanced. String was a lose thread from the sterile towel. The thread had to be removed from around the catheter and before the catheter could be installed and placed successfully. Manufacturer response for sterile disposable towel, sterile disposable towel (per site reporter). They are to make their investigation a priority.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number7489978
MDR Report Key7489978
Date Received2018-05-07
Date of Report2018-05-04
Date of Event2018-04-30
Report Date2018-05-04
Date Reported to FDA2018-05-04
Date Reported to Mfgr2018-05-04
Date Added to Maude2018-05-07
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMEDLINE
Generic NameFIBER, MEDICAL, ABSORBENT
Product CodeFRL
Date Received2018-05-07
Model NumberMDT2168210
Catalog NumberMDT2168210
Lot Number84318010065
ID NumberSTERILE DISPOSABLE TOWEL
OperatorPHYSICIAN
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMEDLINE INDUSTRIES, INC.
Manufacturer Address1204 TOWNLINE RD ONE MEDLINE PL. MUNDELEIN IL 60060 US 60060


Patients

Patient NumberTreatmentOutcomeDate
10 2018-05-07