LINEAR STRAIGHT BROACH HANDLE 803-03-159 510397

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2018-05-07 for LINEAR STRAIGHT BROACH HANDLE 803-03-159 510397 manufactured by Greatbatch Medical.

Event Text Entries

[108141947] The device was returned to integer for evaluation and the reported event was confirmed. The locating pin was broken off of the broach handle and was not returned with the device. It was observed that the locating pin had been welded in place before breaking off. Per print, this locating pin is to be machined as part of the broach handle body, not welded. Manufacturing records were reviewed and there were no discrepancies found which were not already known. Previous to the event date of this complaint, recall (b)(4) had been submitted and the supplier has since been removed from the approved supplier list. This device lot falls within the scope of this recall. No further investigation required. Complaint information provided by distributor, (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[108141948] It was reported during an unknown patient procedure that while broaching, it was noticed a little piece had broken off. There was no surgical delay nor any adverse events reported as a result of this malfunction. The procedure was completed as intended with another device.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004976965-2018-00032
MDR Report Key7490687
Report SourceDISTRIBUTOR
Date Received2018-05-07
Date of Report2018-04-09
Date of Event2018-01-24
Date Mfgr Received2018-04-09
Device Manufacturer Date2016-01-12
Date Added to Maude2018-05-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. TONY SINGH
Manufacturer Street4545 KROEMER ROAD
Manufacturer CityFORT WAYNE IN 46818
Manufacturer CountryUS
Manufacturer Postal46818
Manufacturer Phone2607557466
Manufacturer G1GREATBATCH MEDICAL
Manufacturer Street4545 KROEMER ROAD
Manufacturer CityFORT WAYNE IN 46818
Manufacturer CountryUS
Manufacturer Postal Code46818
Single Use3
Previous Use Code3
Removal Correction NumberZ-2055-2017
Event Type3
Type of Report3

Device Details

Brand NameLINEAR STRAIGHT BROACH HANDLE
Generic NameBROACH HANDLE
Product CodeHTR
Date Received2018-05-07
Returned To Mfg2018-04-11
Model Number803-03-159
Catalog Number510397
Lot Number3185601
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerGREATBATCH MEDICAL
Manufacturer Address4545 KROEMER ROAD FORT WAYNE IN 46818 US 46818


Patients

Patient NumberTreatmentOutcomeDate
10 2018-05-07

© 2026 FDA.report