PROBNP II, N-TERMINAL PRO B-TYPE NATRIURETIC PEPTIDE 04842464190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-05-07 for PROBNP II, N-TERMINAL PRO B-TYPE NATRIURETIC PEPTIDE 04842464190 manufactured by Roche Diagnostics.

Event Text Entries

[107570071] (b)(4). This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[107570072] The customer received questionable elecsys probnp ii immunoassay results for one patient. The erroneous results were reported outside of the laboratory and were believed to be not high enough by the doctor since the patient was diagnosed with acute coronary heart failure (acute chf). The initial result was 338 pg/ml and the repeat result was 335. 4 pg/ml. As analyte-specific antibodies or an interference was suspected, the customer repeated the sample with dilutions. With a 1:2 dilution, the result was 324. 7 pg/ml. With a 1:10 dilution, the result was 335. 4 pg/ml. There was no allegation of an adverse event. The customer used cobas e 411 immunoassay analyzer serial number (b)(4). As no sample could be provided, further investigation was not possible. The investigation was unable to find a definitive root cause.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-01406
MDR Report Key7492031
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-05-07
Date of Report2018-05-07
Date of Event2018-04-20
Date Mfgr Received2018-04-23
Date Added to Maude2018-05-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NamePROBNP II, N-TERMINAL PRO B-TYPE NATRIURETIC PEPTIDE
Generic NameTEST,NATRIURETIC PEPTIDE
Product CodeNBC
Date Received2018-05-07
Model NumberNA
Catalog Number04842464190
Lot Number29280800
ID NumberNA
Device Expiration Date2019-01-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-05-07

© 2024 FDA.report