WARMER DRAPE ORS-300

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2018-05-07 for WARMER DRAPE ORS-300 manufactured by Microtek Dominicana, S.a..

Event Text Entries

[107581203] A sample was received for investigation. The sample drape was placed over a warmer and water was added. The drape was left on the warmer for over 1 hour. No leaks, tears, holes or burn marks were found. The dhr was reviewed for lot d173059 and it was noticed that this lot had 2736 pieces that were manufactured on 11/02/2017. No defects were reported during quality inspections. Based on the device history record and sample review, this does not appear to be the result of a personnel, process, or material issue as no leaks, tears, holes or burn marks were found. Since the reported issue was not found review of the sample and no defects were reported during quality inspection, no additional actions will be taken at this time.
Patient Sequence No: 1, Text Type: N, H10


[107581204] Company rep reported "drape is tearing during cases. Did not specify origins of tearing but that they considered double draping as a solution. I did let them know this is not a good idea and could lead to more issues. " no patient injury or treatment was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8043817-2018-00015
MDR Report Key7492348
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2018-05-07
Date of Report2018-04-09
Date Mfgr Received2018-04-09
Date Added to Maude2018-05-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS JENNIFER MRKVICKA
Manufacturer Street1 ECOLAB PLACE
Manufacturer CityST. PAUL MN 55102
Manufacturer CountryUS
Manufacturer Postal55102
Manufacturer Phone6512502237
Manufacturer G1MICROTEK DOMINICANA
Manufacturer StreetZONA FRANCA NO. 2
Manufacturer CityLA ROMANA,
Manufacturer CountryDR
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWARMER DRAPE
Generic NameFLUID WARMING AND SLUSH DRAPES
Product CodeLHC
Date Received2018-05-07
Returned To Mfg2018-04-27
Model NumberORS-300
Lot NumberD173059
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMICROTEK DOMINICANA, S.A.
Manufacturer AddressZONA FRANCA NO. 2 LA ROMANA, DR


Patients

Patient NumberTreatmentOutcomeDate
10 2018-05-07

© 2024 FDA.report