PENCAN N/A 4502027

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2018-05-08 for PENCAN N/A 4502027 manufactured by B. Braun Melsungen Ag.

Event Text Entries

[107707126] (b)(4). We received one pencan 0,42/88 (g27 x 3 1/2") in open packaging. The returned sample was taken to a visual examination. Pre-existing damages or other faults especially at the raw cannula were not detected on the provided sample. The pencan cannula is broken off approx. 50 mm away from the cannula tip. The structure of the break on the raw cannula and also from the broken-off part shows that the cannula was bent before the break. We assume of problems during application and consider the complaint not confirmed. Device history record (dhr) : unable to review the device history record as it is an unknown batch number. Awaiting sample.
Patient Sequence No: 1, Text Type: N, H10


[107707127] As reported by the user facility ((b)(4)): needle broken.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610825-2018-00078
MDR Report Key7494786
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2018-05-08
Date of Report2018-05-08
Date of Event2018-04-12
Date Facility Aware2018-05-08
Report Date2018-05-08
Date Reported to FDA2018-05-08
Date Reported to Mfgr2018-05-08
Date Mfgr Received2018-04-12
Date Added to Maude2018-05-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JONATHAN SEVERINO
Manufacturer Street861 MARCON BLVD.
Manufacturer CityALLENTOWN 18109
Manufacturer CountryUS
Manufacturer Postal18109
Manufacturer Phone4842408332
Manufacturer G1B. BRAUN MELSUNGEN AG
Manufacturer StreetCARL-BRAUN-STR. 1
Manufacturer CityMELSUNGEN, 34212
Manufacturer CountryGM
Manufacturer Postal Code34212
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePENCAN
Generic NameSPINAL NEEDLE
Product CodeMIA
Date Received2018-05-08
Returned To Mfg2018-04-26
Model NumberN/A
Catalog Number4502027
Lot NumberUNKNOWN
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerB. BRAUN MELSUNGEN AG
Manufacturer AddressCARL-BRAUN-STR. 1 MELSUNGEN, HESSEN 34212 GM 34212


Patients

Patient NumberTreatmentOutcomeDate
10 2018-05-08

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.