MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-05-10 for VITEK? 2 GRAM POSITIVE ID TEST KIT 21342 manufactured by Biomerieux Inc..
[108503110]
A customer from (b)(6) notified biom? Rieux of misidentification results for an rcpa eqa quality control strain when testing with vitek? 2 gp id test kit (ref. 21342). The intended result was abiotrophia defectiva, and vitek 2 identified the organism as granulicatella adiacens. As there is no patient associated with this quality control strain, there is no adverse event related to any patient's state of health. A biom? Rieux internal investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1950204-2018-00183 |
MDR Report Key | 7502227 |
Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
Date Received | 2018-05-10 |
Date of Report | 2018-06-25 |
Date Mfgr Received | 2018-06-05 |
Device Manufacturer Date | 2017-04-01 |
Date Added to Maude | 2018-05-10 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MRS. CANDACE MARTIN |
Manufacturer Street | 595 ANGLUM ROAD |
Manufacturer City | HAZELWOOD MO 63042 |
Manufacturer Country | US |
Manufacturer Postal | 63042 |
Manufacturer Phone | 3147317301 |
Manufacturer G1 | BIOMERIEUX INC. |
Manufacturer Street | 595 ANGLUM ROAD |
Manufacturer City | SAINT LOUIS MO 63042 |
Manufacturer Country | US |
Manufacturer Postal Code | 63042 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | VITEK? 2 GRAM POSITIVE ID TEST KIT |
Generic Name | VITEK? 2 GP ID TEST KIT |
Product Code | LQL |
Date Received | 2018-05-10 |
Catalog Number | 21342 |
Lot Number | 2420309403 |
Device Expiration Date | 2018-10-01 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | BIOMERIEUX INC. |
Manufacturer Address | 595 ANGLUM ROAD SAINT LOUIS MO 63042 US 63042 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2018-05-10 |