[108362591] This event occurred in (b)(6). Unique identifier (udi)#: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[108362592] The customer questioned results for 1 patient tested for elecsys t4 (t4) and elecsys (tsh) on a cobas e801 module. The patient had a thyroid gland resection in 2009 due to struma multinodosa. The results from a thyroid scintigraphy confirm a small amount of residual thyroid tissue. The patient has a family history of thyroid hormone resistance with autosomal dominant inheritance. The patient is treated intravenously with 2mg of l-thyroxine per day. The patient is clinically considered to have hypothyroidism. The customer is concerned that the tsh results are always high and the t4 results are low. This medwatch will cover t4. Refer to medwatch with patient identifier (b)(6) for information on the tsh results. Refer to attached data for the patient results. There was no allegation that an adverse event occurred. The e801 module serial number was (b)(4). The customer stated calibration and qc results were within the acceptable ranges with no issues. The customer is wondering if there is an interference in the patient sample affecting the results.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-01439
MDR Report Key7502569
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-05-10
Date of Report2018-05-10
Date of Event2018-04-23
Date Mfgr Received2018-04-23
Date Added to Maude2018-05-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameELECSYS T4 ASSAY
Generic NameRADIOIMMUNOASSAY, TOTAL THYROXINE
Product CodeCDX
Date Received2018-05-10
Model NumberT4
Catalog Number07027885190
Lot NumberASKU
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-05-10