ATRICURE SYNERGY ABLATION SYSTEM OSL2 A000432

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2018-05-10 for ATRICURE SYNERGY ABLATION SYSTEM OSL2 A000432 manufactured by Atricure, Inc..

Event Text Entries

[107870670] (b)(4). The device was returned and visually and functionally tested. The device met all criteria and functioned normally. The complaint could not be confirmed. There was no report of any device malfunctions.
Patient Sequence No: 1, Text Type: N, H10


[107870671] On (b)(6) 2018 a (b)(6) female patient underwent a on-pump avr/pvi/ascending procedure. The surgeon used an osl2 and performed 4 ablations and cleaned the clamp in between each ablation. At the end of the ablations, the surgeon found a hole near where the clamp had been used near the left pv at the end of the case. There was bleeding that was controlled by suture but did not necessitate a transfusion. Patient was fully heparinized and the procedure was prolonged 20 minutes. Post- procedure patient outcome was good.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3011706110-2018-00161
MDR Report Key7503022
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2018-05-10
Date of Report2018-05-10
Date of Event2018-04-17
Date Mfgr Received2018-04-18
Device Manufacturer Date2016-09-02
Date Added to Maude2018-05-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactANUPAM BEDI
Manufacturer Street7555 INNOVATION WAY
Manufacturer CityMASON OH 45040
Manufacturer CountryUS
Manufacturer Postal45040
Manufacturer G1ATRICURE, INC.
Manufacturer Street7555 INNOVATION WAY
Manufacturer CityMASON OH 45040
Manufacturer CountryUS
Manufacturer Postal Code45040
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameATRICURE SYNERGY ABLATION SYSTEM
Generic NameATRICURE SYNERGY ABLATION SYSTEM
Product CodeOCM
Date Received2018-05-10
Returned To Mfg2018-05-01
Model NumberOSL2
Catalog NumberA000432
Lot Number67706
Device Expiration Date2018-08-01
OperatorOTHER HEALTH CARE PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerATRICURE, INC.
Manufacturer Address7555 INNOVATION WAY MASON OH 45040 US 45040


Patients

Patient NumberTreatmentOutcomeDate
101. Life Threatening; 2. Required No Informationntervention 2018-05-10

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.