FREE PSA 03289788190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-05-10 for FREE PSA 03289788190 manufactured by Roche Diagnostics.

Event Text Entries

[108250446] (b)(4). This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[108250447] The customer experienced an issue where the elecsys free psa immunoassay results were higher than the elecsys total psa immunoassay results for two samples from one patient. The customer used the cobas e 411 immunoassay analyzer serial number (b)(4). For the initial sample, the free psa result was 5. 56 ng/ml and the total psa result was 0. 25 ng/ml. The initial sample was repeated. The free psa result was 5. 77 ng/ml and the total psa result was 0. 23 ng/ml. The initial sample was repeated with an automatic 1:2 dilution. The free psa result was 5. 78 ng/ml and the total psa result was 0. 27 ng/ml. For the initial sample tested with the abbott method, the free psa result was 0. 08 ng/ml and the total psa result was 0. 56 ng/ml. On (b)(6) 2018 a new sample was collected from the patient. The free psa result was 5. 43 ng/ml and the total psa result was 0. 24 ng/ml. No erroneous results were reported outside of the laboratory. There was no allegation of an adverse event. This medwatch will cover the free psa results. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-01444
MDR Report Key7503476
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-05-10
Date of Report2018-06-12
Date of Event2018-04-17
Date Mfgr Received2018-04-24
Date Added to Maude2018-05-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameFREE PSA
Generic NameTEST,PROSTATE SPECIFIC ANTIGEN,FREE,(NONCOMPLEXED) TO DISTINGUISH PROSTATE CA
Product CodeMTG
Date Received2018-05-10
Model NumberNA
Catalog Number03289788190
Lot Number24313902
ID NumberNA
Device Expiration Date2018-09-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-05-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.