ETEST? CEFTAZIDIME TZ 256 WW S30 412293

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-05-10 for ETEST? CEFTAZIDIME TZ 256 WW S30 412293 manufactured by Biomerieux Sa.

Event Text Entries

[108616790] A customer in (b)(6) reported false susceptible ceftazidime (caz) results for a pseudomonas aeruginosa (atcc 27853) quality control strain, and a pseudomonas strain isolated from a patient in intensive care, in association with the etest? Ceftazidime tz 256 ww s30 (lot 1005376560). The customer stated that the caz value for the qc strain was mic= 0. 5mg/l (out of range low) instead of mic= 1-4 mg /l. A new qc strain was tested and the result was the same, mic=0. 5mg/l. The patient strain was tested twice on the impacted lot and the result was caz mic = 4mg/l. The customer also performed the broth micro dilution method and a different i2a ast test on the impacted patient strain, and obtained correlated results of mic= 16 mg/l (resistant result). The customer stated that the false result had a clinical impact on the patient, but no additional information was provided. A biom? Rieux internal investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2018-00065
MDR Report Key7504475
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-05-10
Date of Report2018-08-22
Date Mfgr Received2018-07-26
Device Manufacturer Date2016-12-15
Date Added to Maude2018-05-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. DEBRA BROYLES
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147317301
Manufacturer G1BIOMERIEUX SA
Manufacturer Street3 ROUTE DE PORT MICHAUD LA BALME LES GROTTES ISERE 383
Manufacturer CountryFR
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameETEST? CEFTAZIDIME TZ 256 WW S30
Generic NameETEST? CEFTAZIDIME TZ 256 WW S30
Product CodeJWY
Date Received2018-05-10
Catalog Number412293
Lot Number1005376560
Device Expiration Date2018-12-15
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer Address3 ROUTE DE PORT MICHAUD LA BALME LES GROTTES ISERE 383 FR


Patients

Patient NumberTreatmentOutcomeDate
10 2018-05-10