2008K@HOME MACHINE UNKNOWN-2008K@HOME MACHINE

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2018-05-14 for 2008K@HOME MACHINE UNKNOWN-2008K@HOME MACHINE manufactured by Concord Manufacturing.

Event Text Entries

[108225279] The plant investigation is in process. A supplemental mdr will be submitted upon completion of this activity.
Patient Sequence No: 1, Text Type: N, H10


[108225280] A medical professional (md) reported that when the reverse osmosis (r. O) machine for one of their hemodialysis (hd) patient needed maintenance and was replaced, the patient immediately began having a bad taste in his mouth, headache, became hypertensive, and reported having an edema during treatment. Upon follow up, the md stated that some symptoms (unspecified) went away after the patient stopped using saline flushes during treatment; however the patient symptoms of headache and bad taste still persisted. Per md the hose that connects the r. O machine to the 2008k@ home machine was replaced again and the symptoms returned. The patient is now performing in-center hemodialysis (hd) with no further issues. Per md the cause of patient symptoms is unknown. Additional information regarding patient treatment records, patient demographics, machine information, r. O information was requested; however, not provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2937457-2018-01337
MDR Report Key7512679
Report SourceHEALTH PROFESSIONAL
Date Received2018-05-14
Date of Report2018-05-25
Date of Event2017-12-01
Date Mfgr Received2018-05-16
Date Added to Maude2018-05-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTHOMAS C. JOHNSON
Manufacturer Street920 WINTER ST.
Manufacturer CityWALTHAM MA 02451
Manufacturer CountryUS
Manufacturer Postal02451
Manufacturer Phone7816999499
Manufacturer G1CONCORD MANUFACTURING
Manufacturer Street4040 NELSON AVENUE
Manufacturer CityCONCORD CA 94520
Manufacturer CountryUS
Manufacturer Postal Code94520
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name2008K@HOME MACHINE
Generic NameHEMODIALYSIS SYSTEM FOR HOME USE
Product CodeONW
Date Received2018-05-14
Catalog NumberUNKNOWN-2008K@HOME MACHINE
OperatorLAY USER/PATIENT
Device AvailabilityN
Device AgeMO
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCONCORD MANUFACTURING
Manufacturer Address4040 NELSON AVENUE CONCORD CA 94520 US 94520


Patients

Patient NumberTreatmentOutcomeDate
10 2018-05-14

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.