APOLLOBLUE 0007567-0015

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-05-15 for APOLLOBLUE 0007567-0015 manufactured by Lap Laser Gmbh.

Event Text Entries

[108807100] A service engineer in the (b)(6) cancer treatment centre, (b)(6), went to unplug the power supply unit (psu) of the medical device apolloblue and the entire casing of the psu came away in his hand, exposing potentially live parts inside. This represents a health and safety hazard which might result in electrical shock. The psu is of type fw7660m/24 of the company friwo geratebau (b)(4). Used batch is 47130. There has been no harm or serious deterioration of persons.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9681134-2018-04171
MDR Report Key7514835
Date Received2018-05-15
Date of Report2018-04-17
Date of Event2018-02-13
Date Mfgr Received2018-02-20
Device Manufacturer Date2018-05-12
Date Added to Maude2018-05-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationHOSPITAL SERVICE TECHNICIAN
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. HANS-PETER WERNER
Manufacturer StreetZEPPELINSTRASSE 23
Manufacturer CityLUNEBURG, 21337
Manufacturer CountryGM
Manufacturer Postal21337
Manufacturer G1LAP LASER GMBH
Manufacturer StreetZEPPELINSTRASSE 23
Manufacturer CityLUNEBURG, 21337
Manufacturer CountryGM
Manufacturer Postal Code21337
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameAPOLLOBLUE
Generic NameLASER FOR PATIENT ALIGNMENT
Product CodeIWE
Date Received2018-05-15
Returned To Mfg2018-03-06
Model Number0007567-0015
OperatorHOSPITAL SERVICE TECHNICIAN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerLAP LASER GMBH
Manufacturer AddressZEPPELINSTRASSE 23 LUNEBURG, 21337 GM 21337


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2018-05-15